279 FC
Calibration Manual
12
Table 3. Performance Tests
Test
(Control Knob Position)
Input
Lower Limit
Upper Limit
Units
Volts AC
5 V, 45 Hz
4.947
5.053
V ac
5 V, 500 Hz
4.247
5.753
3 V, 45 Hz
2.967
3.033
50 V, 45 Hz
49.47
50.53
50 V, 500 Hz
42.47
57.53
30 V, 45 Hz
29.67
30.33
500 V, 45 Hz
494.7
505.3
500 V, 500 Hz
424.7
575.3
1000 V, 45 Hz
987
1013
Volts AC, Frequency
1 V, 900 Hz
899.0
901.0
Hz
5 V, 50 Hz
49.94
50.06
DC Volts
0.01 V
0.008
0.012
V dc
-5 V
-5.006
-4.994
50 V
49.94
50.06
10 V
9.97
10.03
-500 V
-500.6
-499.4
1000 V
997
1003
DC Millivolts
3 mV
2.8
3.2
mV dc
500 mV
499.4
500.6
-500 mV
-500.6
-499.4
AC Millivolts
30 mV, 60 Hz
29.4
30.6
mV ac
500 mV, 60 Hz
494.7
505.3
500 mV, 500 Hz
494.7
505.3
Ohms
6
Ω
5.8
6.2
Ω
500
Ω
497.3
502.7
0.6 k
Ω
0.596
0.604
k
Ω
5 k
Ω
4.974
5.026
50 k
Ω
49.74
50.26
500 k
Ω
497.4
502.6
5 M
Ω
4.974
5.026
M
Ω
50 M
Ω
49.22
50.78
Capacitance
10 nF
8
12
nF
900 nF
887
913
9
µ
F
8.87
9.13
µ
F
90
µ
F
88.7
91.3
900
µ
F
887
913
Continuity
24
Ω
Beeper On
251
Ω
Beeper Off
Diode
1.95 V
1.928
1.972
V dc
Simulation
3.6 mV, 60 Hz
96.5
103.5
A ac
36 mV, 60 Hz
965.0
1035
120 mV, 100 Hz
1940
2060
IR CAMERA
-5
°
C
-10
0
°
C
50
°
C
+45
+55
180
°
C
+171
+189
Содержание 279 FC