393/393 FC
Users Manual
20
AC Voltage
Range .............................................................. 600.0 V
1000 V
Resolution........................................................ 0.1 V (
≤
600.0 V)
1 V (
≤
1000 V)
Accuracy.......................................................... 1 % RD
+
5 digits (20 Hz to 500 Hz)
DC Voltage
Range .............................................................. 600.0 V
1500 V
Resolution........................................................ 0.1 V (
≤
600.0 V)
1 V (
≤
1500 V)
Accuracy.......................................................... 1 % RD
+
5 digits
mV dc
Range .............................................................. 500.0 mV
Resolution........................................................ 0.1 mV
Accuracy.......................................................... 1 % RD
+
5 digits
Amps Frequency: Jaw
Range .............................................................. 5.0 Hz to 500.0 Hz
Resolution........................................................ 0.1 Hz
Accuracy.......................................................... 0.5 % RD
+
5 digits
Trigger Level.................................................... 5 Hz to 10 Hz,
≥
10 A
10 Hz to 100 Hz,
≥
5 A
100 Hz to 500 Hz,
≥
10 A
Amps Frequency: Flexible Current Probe
Range .............................................................. 5.0 Hz to 500.0 Hz
Resolution........................................................ 0.1 Hz
Accuracy.......................................................... 0.5 % RD
+
5 digits
Trigger Level.................................................... 5 Hz to 20 Hz,
≥
25 A
20 Hz to 100 Hz,
≥
20 A
100 Hz to 500 Hz,
≥
25 A
Voltage Frequency
Range .............................................................. 5.0 Hz to 500.0 Hz
Resolution........................................................ 0.1 Hz
Accuracy.......................................................... 0.5 % RD
+
5 digits
Trigger Level.................................................... 5 Hz to 20 Hz,
≥
5 V
20 Hz to 100 Hz,
≥
5 V
100 Hz to 500 Hz,
≥
10 V