CAT III 1500V TRMS Clamp Meter
Measurement Features
15
2. When in Auto Hold mode, push
e
>2 s again to exit the Auto Hold mode.
Note
Auto Power Off is always disabled when you use the Auto Hold function. When Auto Hold
is enabled on VA, push
to disable it automatically. Auto hold is disabled when
Inrush Current/Peak Current or Min/Max/Ave is enabled.
Min/Max/Avg Measurements
Min/Max/Avg mode captures the minimum, maximum, and average readings of a given output
signal over an extended time. The Clamp beeps when it senses a new high value or new low
value. It applies for both reading except for Inrush Current/Peak Current. Push
e
to
pause the
reading update (recording continues).
This function works in current, voltage, and frequency modes:
1. Push
M
to enter the Min/Max/Avg mode.
The maximum reading shows on the display.
2. Continue to push
M
to select between the maximum, minimum, average, and live readings.
The cycle continues each time you push
M
.
3. To exit Min/Max/Avg mode, push and hold
M
for >2 s.
Note
The Min/Max/Avg function does not support Auto Hold and Inrush Current/Peak Current.
Auto Power Off is always disabled when you use the Min/Max/Avg, Auto Hold, and
Logging functions or Fluke Connect is active.
Table 5. Auto Hold Functions
Function
Threshold
Fluctuation Range/Delta
V ac
10 V
2 V
V dc
10 V
2 V
mV dc
20 mV
5 mV
Ohm
60 k
Ω
2.0
Ω
/20
Ω
/0.20 k
Ω
Capacitance
10
μ
F
2
μ
F
A dc
10 A
2 A
A ac
10 A for Clamp/25 A for iFlex
2 A for Clamp/5 A for iFlex
dc Power
10.0 kVA
2.0 kVA