CAT III 1500V TRMS Clamp Meter
Basic Measurements
11
Amps DC
To measure dc current:
1. Turn control knob to
}
.
2. Push
to compensate (zero) for outside influences.
3. Position the Clamp jaw around the conductor.
The display shows the value and
C
to indicate that the measurement is from the jaw. When
the current measurement is <0.5 A, the center dot in the icon flashes. For current
measurements >0.5 A, the center dot in the icon is steady.
Power DC
To measure dc power:
1. Turn control knob to
}
.
2. Push
to compensate (zero) for outside influences.
3. Turn control knob to
{
.
4. Position the Clamp jaw around the conductor.
5. Connect the black test lead to the
COM
terminal and the red test lead to the
V
X
terminal.
6. Touch the probes to the test points of the circuit.
The display shows the measurement of dc power and dc current.
The display also shows
C
to indicate that the measurement is from the jaw.
Note
Push
to toggle the readout between dc power and dc voltage.