This group of commands deals with the selection of input channels. These
commands are summarized in Table 12.
6.5.3.1
ROUTe:CLOSe <channel>
This command selects the primary input channel. It is equivalent to setting the
primary channel using the
PRIM CHAN
soft-key function in the
INPUT
menu. The <channel> parameter specifies the channel number. The channel
number is preceded by @ and enclosed in parentheses. This command also sets
the scan state to OFF (see Section 6.5.3.8) and the scanning alternate state to
OFF (see Section 6.5.3.4). It does not affect the measure mode or INITiate
state. The *RST command sets the primary channel to channel 1.
Example command: ROUT:CLOS (@3)
6.5.3.2
ROUTe:CLOSe:STATe?
This query command returns the number of the current input channel.
Example command: ROUT:CLOS:STAT?
Example response: 3
6.5.3.3
ROUTe:PRIMary?
This query command returns the number of the primary channel.
Example command: ROUT:PRIM?
Example response: 1
1560 Thermometer Readout
User’s Guide
78
Command
Action
ROUTe:CLOSe <channel>
Select the primary channel
ROUTe:CLOSe:STATe?
Query the current channel number being measured
ROUTe:PRIMary?
Query primary channel number
ROUTe:SCAN:ALTernate <Boolean>
Set the scan/primary channel alternate mode
ROUTe:SCAN:ALTernate?
Query the alternate mode
ROUTe:SCAN[:LIST] <channel_list>
Select channels for scanning
ROUTe:SCAN[:LIST]?
Query scanning channels
ROUTe:SCAN:STATe <Boolean>
Enable or disable scanning
ROUTe:SCAN:STATe?
Query scanning
Table 12
Input Channel Commands
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...