P5515, P5515-V, & P5515-EP
Users Manual
16
Table 7. Reservoir Assembly - Parts List
Part
Item Description
P5515 P5515-V
P5515-EP
1 VALVE
STEM
3886430
2 SPRING
3922786
3 NYLON
WASHER
3916458
4 RESERVOIR
COVER 3921624
5 RESERVOIR
CAP
3921613
6 O
RING
3864910
3865326
3865552
7 RESERVOIR
TUBE
3921608 3921726 3921726
8 SCREW
3909292
9 WASHER
3205770
10 VALVE
VENT
3921582
11 O
RING
3865142
3865163
3867888
12 RESERVOIR
BODY
3921575
13 ELBOW
3926855
14 PLUG
3068437
15 GLAND
NUT
3903089
Fault Finding
System Will Not Prime
1.
Check reservoir valve is closed.
2.
Check for sufficient fluid in reservoir.
3.
Check for damaged/missing/dirty O-ring on test ports.
4.
Check that the face of the DUT is contacting the O-ring, and that the surface is not
scored or dented.
System Will Not Pressurize
1.
Ensure correct valve operation during priming process.
2.
Check DUT is not leaking.
3.
Clean system externally, check for fluid leak by continually trying to pressurize.
Wherever fluid appears, replace the seal — check sealing faces are clean and
undamaged before re-assembly.
4.
Check that the face of the DUT is contacting the O-ring, and that the surface is not
scored or dented.