Hydraulic Comparison Test Pump
Test Port Insert
7
Test Port Insert
For devices with 1/8 BSP or NPT mounting threads, the diameter of the thread is very
close to the effective sealing diameter of the O-ring fitted to the test port.
This can make it difficult to achieve a good seal. When mounting these devices, use the
test port insert (stored in the spare seals container) as shown in Figure 7.
1
2
3
4
gjn012.eps
Figure 7. Test Port Insert
Table 3. Test Port Insert - Parts
Part
Item
Description
P5515 P5515-V
P5515-EP
1 O
Ring
3865142
3865163 3867888
2
Test Port Insert
3919892
3 O
Ring
3883397
3883521 3867895
4 Test
Port
3921414
To calibrate panel-mounted gauges with pressure connections in the rear, use an Angle
Adapter (see Chapter 8, Ancillary Equipment).
Priming
1.
Open reservoir valve one turn counter-clockwise and turn capstan fully in.
2.
Pump the priming pump two times.
3.
Close valve (see caution below) and turn capstan fully out.
4.
Open valve and turn capstan fully in.