Sr=2.0
New scan rate
Press “SET” to accept the new scan rate and continue.
9.5
Temperature Scale Units
The temperature scale units of the controller may be set by the user to degrees
Celsius (°C) or Fahrenheit (°F). The units are used in displaying the bath tem-
perature, set-point, vernier, proportional band, and cutout set-point.
The temperature scale units selection is accessed after the vernier adjustment
function by pressing “SET”. From the temperature display function access the
units selection by pressing “SET” 4 times.
25.00 C
Bath temperature
S
Access set-point memory
1. 25.0
Set-point memory
S
Access set-point value
C 25.00
Set-point value
S
Access vernier
0.00000
Vernier setting
S
Access scale units selection
Un= C
Scale units currently selected
Press “UP” or “DOWN” to change the units.
U
Change units
Un= F
New units selected
Press “SET” to accept the new selection and resume displaying the bath
temperature.
6331 Calibration Bath
User’s Guide