U
Increment display
0.00090
New vernier setting
Next press “EXIT” to return to the temperature display or “SET” to access the
temperature scale units selection.
S
Access scale units
9.4
Scan
The scan rate can be set and enabled so that when the set-point is changed the
instrument heats or cools at a specified rate (degrees per minute) until it reaches
the new set-point. With the scan disabled the instrument heats or cools at the
maximum possible rate.
9.4.1
Scan Control
The scan is controlled with the scan on/off function that appears in the main
menu after the set-point function.
ScAn=OFF
Scan function off
Press “UP” or “DOWN” to toggle the scan on or off.
ScAn=On
Scan function on
Press “SET” to accept the present setting and continue.
S
Accept scan setting
9.4.2
Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set
from 0.001 to 5.0 °C/min. The maximum scan rate however is actually limited
by the natural heating or cooling rate of the instrument. This is typically less
than 2.5°C/min, especially when cooling. The scan rate function appears in the
main menu after the scan control function. The scan rate units are in degrees
per minute, degrees C or F depending on the selected units.
Sr=0.010
Scan rate in °C/min
Press “UP” or “DOWN” to change the scan rate.
41
9 Controller Operation
Scan