5730A
Operators Manual
4-4
Warmup
Make sure the 5730A Calibrator is warmed up before use. The environmentally-
controlled components inside must be stable to meet or exceed the specifications in
Chapter 1. Sufficient warmup times are:
•
If the Calibrator has been powered off for 1 hour or more, allow at least 30 minutes
of warmup time.
•
If the Calibrator has been powered off for a length of time less than 1 hour, allow it to
warm up for at least twice the length of time it was turned off. For example, if it has
been powered off for 10 minutes, allow at least 20 minutes of warmup time.
Product Use
Touchscreen navigation is used to move throughout the 5730A Calibrator user interface
(UI) and menus.
Reset the Calibrator
At any time during front-panel operation, return the Calibrator to the power-up state with
. When
is pushed, it:
•
Returns the Calibrator to the power-up state: 0 mV dc, standby, internal guard, and
internal sense.
•
Clears the stored values for limits, offset, scale, and error mode reference.
•
Secures the Calibrator if the passcode had been entered.
Operate and Standby Modes
When the OPERATE light above the output terminals is illuminated, the output value and
function shown on the display are active at the selected terminals. “Operate” also shows
on the display.
When the STANDBY light above the output terminals is lit, all 5730A Calibrator binding
posts except GROUND are open-circuited. “Standby” also appears on the display in this
case.
Visual Connection Management output terminals illuminate in green to indicate which
terminals are active and provide guidance to ensure the proper connection of cables for
each function.
Push
to put the Calibrator into the OPERATE state. Push
to put the
Calibrator in the STANDBY state.
If any of the subsequent events occur during operation, the Calibrator automatically
switches to standby mode:
•
The
key is pushed.
•
A voltage
≥
22 V is selected when the previous output voltage was
<
22 V.
•
The output location is changed, for example by selecting an amplifier. The
exception is when the 5725A is selected for ac voltage or current, provided the
Calibrator current output location is set to “5725A.”
•
The output function is changed from one function to another.
Содержание 5730A
Страница 4: ......
Страница 10: ...5730A Operators Manual x ...
Страница 12: ...5730A Operators Manual xii ...
Страница 14: ...5730A Operators Manual xiv ...
Страница 16: ...5730A Operators Manual 1 2 ...
Страница 17: ...Introduction and Specifications Introduction 1 1 3 ...
Страница 18: ......
Страница 52: ...5730A Operators Manual 2 2 ...
Страница 57: ...3 1 Chapter 3 Features Title Page Introduction 3 3 Front Panel Features 3 3 Rear Panel Features 3 7 ...
Страница 58: ...5730A Operators Manual 3 2 ...
Страница 124: ...5730A Operators Manual 5 2 ...
Страница 136: ...5730A Operators Manual 6 2 ...
Страница 198: ...5730A Operators Manual 7 6 115V 11 10 4 5 6 7 8 9 3 2 1 hhp224 eps Figure 7 3 Replaceable Parts ...
Страница 216: ...5730A Operators Manual 7 24 ...
Страница 218: ...5730A Operators Manual 8 2 ...
Страница 222: ...5730A Operators Manual 8 6 ...
Страница 248: ...5730A Operators Manual A 26 ...
Страница 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...