5730A
Operators Manual
6-52
Check the Calibrator Status
The controller has access to six status registers for the 5730A Calibrator, which indicate
the Calibrator conditions in the as shown in Figure 6-1. Each register bit is explained
under separate headings for each register. Table 6-9 lists each register and its remote
commands.
In addition to the status registers, the Service Request control line, SRQ (available only
when an IEEE-488 controller is used) and a 16-element buffer called the Fault Queue
provide status information.
Table 6-9. Status Register Summary
Register READ
Command
WRITE
Command
Enable
Register
Serial Poll Status Byte (STB)
*STB? (or SPL( )
for some controllers)
None SRE
Service Request Enable Register
(SRE)
*SRE? *SRE
None
Event Status Register (ESR)
*ESR?
None
ESE
Event Status Enable Register (ESE)
*ESE?
*ESE
None
Instrument Status Register (ISR)
ISR?
None
None
Instrument Status Change Register
(ISCR)
ISCR? None
ISCE
Instrument Status Change Enable
Register (ISCE)
ISCE? ISCE
None
Содержание 5730A
Страница 4: ......
Страница 10: ...5730A Operators Manual x ...
Страница 12: ...5730A Operators Manual xii ...
Страница 14: ...5730A Operators Manual xiv ...
Страница 16: ...5730A Operators Manual 1 2 ...
Страница 17: ...Introduction and Specifications Introduction 1 1 3 ...
Страница 18: ......
Страница 52: ...5730A Operators Manual 2 2 ...
Страница 57: ...3 1 Chapter 3 Features Title Page Introduction 3 3 Front Panel Features 3 3 Rear Panel Features 3 7 ...
Страница 58: ...5730A Operators Manual 3 2 ...
Страница 124: ...5730A Operators Manual 5 2 ...
Страница 136: ...5730A Operators Manual 6 2 ...
Страница 198: ...5730A Operators Manual 7 6 115V 11 10 4 5 6 7 8 9 3 2 1 hhp224 eps Figure 7 3 Replaceable Parts ...
Страница 216: ...5730A Operators Manual 7 24 ...
Страница 218: ...5730A Operators Manual 8 2 ...
Страница 222: ...5730A Operators Manual 8 6 ...
Страница 248: ...5730A Operators Manual A 26 ...
Страница 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...