Introduction
and
Specifications
Support Equipment and Services
1
1-11
5725A Amplifier
The Fluke Calibration 5725A Amplifier is an external unit that operates under calibrator
control. It extends ac voltage drive capabilities and both ac and dc current output range.
The amplifier adds these capabilities to the 1100 V ac range of the 5730A Calibrator with
no compromise in accuracy:
•
Frequency limits at higher voltage increase to 100 kHz at 750 V, 30 kHz at 1100 V.
•
Load limit increases to 70 mA for frequencies above 5 kHz.
•
Capacitive drive increases to 1000 pF, subject to the maximum output current.
A separate set of binding posts on the front panel of the 5725A supplies extended-range
ac and dc current outputs. Since most meters have a separate input terminal for the high-
current ranges, this eliminates the need to change cables during a procedure. The 5725A
can also be configured to source all current (both standard calibrator-generated current
and its own current) through the 5725A binding posts.
52120A Amplifier
The Fluke Calibration 52120A Transconductance Amplifier is an external unit that
operates under calibrator control to extend the ac and dc current output range of the
5730A Calibrator. A maximum of three 52120A amplifiers can be connected, as much as
tripling the current output available. The 52120A amplifier can:
•
Accept full scale dc or ac inputs of 2 volts or 200 mA from any calibrator, signal
generator or power supply
•
Deliver proportional output current in ranges of 2 A, 20 A or 120 A at frequencies to
10 kHz
•
Offer enhanced accuracy to 140 ppm when used in closed-loop mode with a 6105A
Electrical Power Standard
•
Operate in parallel with one or two other 52120As to deliver 240 A or 360 A
•
Source current with compliance voltage of 4.5 V rms or 6.4 V peak
•
Drive inductive loads to 1 mH
•
Drive optional current coils to deliver test currents of 3000 A or 6000 A
Support Equipment and Services
Fluke Calibration supports calibration requirements with precision, high-quality
equipment and a wide range of services. Depending on the calibration needs, location,
and capabilities, the 5730A Calibrator can be supported independently or with Fluke
Calibration services for part, or all support needs. The subsequent paragraphs describe the
support equipment and services offered by Fluke Calibration for the Calibrator. For
specifications and ordering instructions for this support equipment and other Fluke
Calibration instruments, refer to the Fluke Calibration catalog, or contact a representative
at a Fluke Calibration Sales and Service Center. See “How to Contact Fluke Calibration”.
Содержание 5730A
Страница 4: ......
Страница 10: ...5730A Operators Manual x ...
Страница 12: ...5730A Operators Manual xii ...
Страница 14: ...5730A Operators Manual xiv ...
Страница 16: ...5730A Operators Manual 1 2 ...
Страница 17: ...Introduction and Specifications Introduction 1 1 3 ...
Страница 18: ......
Страница 52: ...5730A Operators Manual 2 2 ...
Страница 57: ...3 1 Chapter 3 Features Title Page Introduction 3 3 Front Panel Features 3 3 Rear Panel Features 3 7 ...
Страница 58: ...5730A Operators Manual 3 2 ...
Страница 124: ...5730A Operators Manual 5 2 ...
Страница 136: ...5730A Operators Manual 6 2 ...
Страница 198: ...5730A Operators Manual 7 6 115V 11 10 4 5 6 7 8 9 3 2 1 hhp224 eps Figure 7 3 Replaceable Parts ...
Страница 216: ...5730A Operators Manual 7 24 ...
Страница 218: ...5730A Operators Manual 8 2 ...
Страница 222: ...5730A Operators Manual 8 6 ...
Страница 248: ...5730A Operators Manual A 26 ...
Страница 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...