5730A AC Current Specifications: 95 % Confidence Level
Range
Resolution
Frequency
(Hz)
Absolute Accuracy
±
5
°
C from calibration
temperature
Relative Accuracy
±
1
°
C
24 Hours
90 Days
180 Days
1 Year
24 Hours
90 Days
±
(ppm nA)
220
μ
A
1 nA
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
210 + 16
130 + 10
96 + 8
240 + 12
800 + 65
230 + 16
140 + 10
99 + 8
250 + 12
900 + 65
240 + 16
150 + 10
101 + 8
270 + 12
1000 + 65
250 + 16
160 + 10
103 + 8
280 + 12
1100 + 65
210 + 16
110 + 10
80 + 8
200 + 12
700 + 65
230 + 16
130 + 10
90 + 8
230 + 12
800 + 65
2.2 mA
10 nA
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
210 + 40
130 + 35
96 + 35
170 + 110
800 + 650
230 + 40
140 + 35
99 + 35
180 + 110
900 + 650
240 + 40
150 + 35
101 + 35
190 + 110
1000 + 650
250 + 40
160 + 35
103 + 35
200 + 110
1100 + 650
210 + 40
110 + 35
80 + 35
160 + 110
700 + 650
230 + 40
130 + 35
90 + 35
170 + 110
800 + 650
22 mA
100 nA
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
210 + 400
130 + 350
96 + 350
170 + 550
800 + 5000
230 + 400
140 + 350
99 + 350
180 + 550
900 + 5000
240 + 400
150 + 350
101 + 350
190 + 550
1000 + 5000
250 + 400
160 + 350
103 + 350
200 + 550
1100 + 5000
210 + 400
110 + 350
80 + 350
160 + 550
700 + 5000
230 + 400
130 + 350
90 + 350
170 + 550
800 + 5000
±
(ppm
μ
A)
220 mA
1
μ
A
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
210 + 4
130 + 3.5
96 + 2.5
170 + 3.5
800 + 10
230 + 4
140 + 3.5
99 + 2.5
180 + 3.5
900 + 10
240 + 4
150 + 3.5
101 + 2.5
190 + 3.5
1000 + 10
250 + 4
160 + 3.5
103 + 2.5
200 + 3.5
1100 + 10
210 + 4
110 + 3.5
80 + 2.5
160 + 3.5
700 + 10
230 + 4
130 + 3.5
90 + 2.5
170 + 3.5
800 + 10
2.2 A
10
μ
A
20 - 1 k
1 k - 5 k
5 k - 10 k
214 + 35
350 + 80
5000 + 160
224 + 35
390 + 80
6000 + 160
234 + 35
420 + 80
6500 + 160
244 + 35
450 + 80
7000 + 160
200 + 35
300 + 80
5000 + 160
230 + 35
350 + 80
6000 +
160
5725A Amplifier:
11 A
100
μ
A
40 - 1 k
1 k - 5 k
5 k - 10 k
370 + 170
800 + 380
3000 + 750
400 + 170
850 + 380
3300 + 750
440 + 170
900 + 380
3500 + 750
460 + 170
950 + 380
3600 + 750
300 + 170
700 + 380
2800 + 750
330 + 170
800 + 38
3200 +
750
Note:
Maximum output from the calibrator’s terminals is 2.2 A. Accuracy specifications for 220
μ
A and 2.2 mA ranges are increased by 1.3
plus 2
μ
A when supplied through 5725A terminals.
Содержание 5730A
Страница 4: ......
Страница 10: ...5730A Operators Manual x ...
Страница 12: ...5730A Operators Manual xii ...
Страница 14: ...5730A Operators Manual xiv ...
Страница 16: ...5730A Operators Manual 1 2 ...
Страница 17: ...Introduction and Specifications Introduction 1 1 3 ...
Страница 18: ......
Страница 52: ...5730A Operators Manual 2 2 ...
Страница 57: ...3 1 Chapter 3 Features Title Page Introduction 3 3 Front Panel Features 3 3 Rear Panel Features 3 7 ...
Страница 58: ...5730A Operators Manual 3 2 ...
Страница 124: ...5730A Operators Manual 5 2 ...
Страница 136: ...5730A Operators Manual 6 2 ...
Страница 198: ...5730A Operators Manual 7 6 115V 11 10 4 5 6 7 8 9 3 2 1 hhp224 eps Figure 7 3 Replaceable Parts ...
Страница 216: ...5730A Operators Manual 7 24 ...
Страница 218: ...5730A Operators Manual 8 2 ...
Страница 222: ...5730A Operators Manual 8 6 ...
Страница 248: ...5730A Operators Manual A 26 ...
Страница 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...