ESA612
Users Manual
48
Performing a Direct Applied Part Leakage Test
Note
The Direct Applied Part Leakage test is available
when the EN62353 standard is selected.
The Direct Applied Part Leakage Current test measures
the leakage current between all applied parts of a single
function and the exposed conductive surface on the
housing, to mains earth. For equipment with multiple
applied parts, each group of a single function should be
tested each in turn with all other floating during the test.
This test should only be done for equipment with F-Type
applied parts.
For Type B applied part, see direct equipment leakage
schematic in Figure 27.
To perform a direct applied part leakage test:
1. Press
.
2. Press the soft key labeled
More
.
3. Select the desired applied part groupings using
and
.
4. Press the soft key labeled
Select
. The Direct A.P. test
should already be selected.
5. Press
or
to select the applied part test
configuration.
6. Press
to apply the test voltage and read the
current in the display.
7. Press
or
to advance to the next group of applied
parts, if applicable.
Figure 28 shows the electrical connections between the
Analyzer and the DUT during a Direct Applied Part
Leakage Current Test.
The following outlet conditions apply when performing this
test:
•
Normal Polarity
•
Reversed Polarity
Note
If there are more than five applied parts to
connect to the Analyzer, see Using the 1-to-10
Adapter later in this manual.
Содержание ESA612
Страница 6: ......
Страница 10: ...ESA612 Users Manual iv ...
Страница 12: ...ESA612 Users Manual vi ...
Страница 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Страница 34: ...ESA612 Users Manual 20 fis112 png Figure 10 Ground Wire Protective Earth Resistance Measurement Connections ...
Страница 36: ...ESA612 Users Manual 22 faw26 png Figure 11 Ground Wire Protective Earth Resistance Measurement Schematic ...
Страница 38: ...ESA612 Users Manual 24 faw17 png Figure 13 Mains to Protective Earth Insulation Resistance Test Schematic ...
Страница 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Страница 42: ...ESA612 Users Manual 28 faw21 png Figure 17 Applied Parts to Non Earth Conductive Points Schematic ...
Страница 50: ...ESA612 Users Manual 36 gtv29 png Figure 21 Lead to Ground Patient Leakage Current Test Schematic ...
Страница 52: ...ESA612 Users Manual 38 gtv30 png Figure 23 Lead to Lead Patient Auxiliary Leakage Current Test Schematic ...
Страница 58: ...ESA612 Users Manual 44 faw22 png Figure 25 Alternative Equipment Leakage Current Test Schematic ...
Страница 64: ...ESA612 Users Manual 50 gtv25 png Figure 28 Direct Applied Parts Leakage Current Test Schematic ...
Страница 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Страница 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Страница 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Страница 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Страница 88: ......