Electrical Safety Analyzer
Performing Electrical Safety Tests
39
The following outlet conditions apply when performing this
test:
•
Normal Polarity
•
Normal Polarity, Open Neutral
•
Normal Polarity, Open Earth
•
Reversed Polarity, Open Neutral
•
Reversed Polarity, Open Earth
Performing a Lead Isolation (Mains on Applied
Part) Leakage Test
Note
The Lead Isolation (Mains on Applied Part)
leakage test is available when the IEC60601 &
ANSI/AAMI standard is selected.
The Lead Isolation (Mains On Applied Parts) Leakage
Current test measures the current that flows in response
to an isolated AC voltage applied between a selected
applied part, group of applied parts, or ALL applied parts,
and Earth (and any conductive part connected to the RED
terminal). Figure 24 shows the electrical connections
between the Analyzer and the DUT during a Mains on
Applied Part Leakage Current Test.
Note
With 60601 standard selected, the MAP test
voltage is available in both Normal and Reverse
(180 degrees out of phase with mains).
To perform a Lead Isolation (Mains on Applied Part) test:
1. Press
.
2. Press the soft key labeled
More
.
3. Select the desired applied part groupings using
and
.
Note
Refer to the testing standard when deciding the
type of the applied parts and how they should be
grouped for testing.
4. Press the soft key labeled
Select
.
5. Press the soft key labeled
Lead Isolation
.
6. Press
or
to select the desired applied part
connection.
7. Press
to apply the voltage and read the leakage
current in the display.
Pressing
and
scrolls through the applied part
connections or groupings. Press
for each
connection configuration to thoroughly test the DUT.
Содержание ESA612
Страница 6: ......
Страница 10: ...ESA612 Users Manual iv ...
Страница 12: ...ESA612 Users Manual vi ...
Страница 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Страница 34: ...ESA612 Users Manual 20 fis112 png Figure 10 Ground Wire Protective Earth Resistance Measurement Connections ...
Страница 36: ...ESA612 Users Manual 22 faw26 png Figure 11 Ground Wire Protective Earth Resistance Measurement Schematic ...
Страница 38: ...ESA612 Users Manual 24 faw17 png Figure 13 Mains to Protective Earth Insulation Resistance Test Schematic ...
Страница 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Страница 42: ...ESA612 Users Manual 28 faw21 png Figure 17 Applied Parts to Non Earth Conductive Points Schematic ...
Страница 50: ...ESA612 Users Manual 36 gtv29 png Figure 21 Lead to Ground Patient Leakage Current Test Schematic ...
Страница 52: ...ESA612 Users Manual 38 gtv30 png Figure 23 Lead to Lead Patient Auxiliary Leakage Current Test Schematic ...
Страница 58: ...ESA612 Users Manual 44 faw22 png Figure 25 Alternative Equipment Leakage Current Test Schematic ...
Страница 64: ...ESA612 Users Manual 50 gtv25 png Figure 28 Direct Applied Parts Leakage Current Test Schematic ...
Страница 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Страница 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Страница 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Страница 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Страница 88: ......