ESA612
Users Manual
10
How to Hold the Product
When you move the Analyzer, use the handle in the
bottom case to hold it. See Figure 3.
Note
There is no handle on old versions of the
Analyzer.
fis122.png
Figure 3. Product Handle
Connecting to Line Power
Warning
To avoid shock hazard and for proper
Analyzer operation, connect the factory
supplied three-conductor line power cord to
a properly grounded power outlet. Do not
use a two-conductor adapter or extension
cord; this will break the protective ground
connection.
Connect the Analyzer to a properly grounded three-prong
outlet. The Analyzer will not properly test a DUT when the
ground lead is open.
The Analyzer is intended for use with single-phase,
grounded power. It is not intended for dual, split-phase or
three-phase power configurations. But it can be used with
any power system that supplies the correct voltages for
single-phase and is grounded, or is an isolated power
system.
Содержание ESA612
Страница 6: ......
Страница 10: ...ESA612 Users Manual iv ...
Страница 12: ...ESA612 Users Manual vi ...
Страница 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Страница 34: ...ESA612 Users Manual 20 fis112 png Figure 10 Ground Wire Protective Earth Resistance Measurement Connections ...
Страница 36: ...ESA612 Users Manual 22 faw26 png Figure 11 Ground Wire Protective Earth Resistance Measurement Schematic ...
Страница 38: ...ESA612 Users Manual 24 faw17 png Figure 13 Mains to Protective Earth Insulation Resistance Test Schematic ...
Страница 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Страница 42: ...ESA612 Users Manual 28 faw21 png Figure 17 Applied Parts to Non Earth Conductive Points Schematic ...
Страница 50: ...ESA612 Users Manual 36 gtv29 png Figure 21 Lead to Ground Patient Leakage Current Test Schematic ...
Страница 52: ...ESA612 Users Manual 38 gtv30 png Figure 23 Lead to Lead Patient Auxiliary Leakage Current Test Schematic ...
Страница 58: ...ESA612 Users Manual 44 faw22 png Figure 25 Alternative Equipment Leakage Current Test Schematic ...
Страница 64: ...ESA612 Users Manual 50 gtv25 png Figure 28 Direct Applied Parts Leakage Current Test Schematic ...
Страница 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Страница 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Страница 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Страница 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Страница 88: ......