Fluke Biomedical DPM1B Скачать руководство пользователя страница 24

DPM1B 

Operators Manual 

14 

Maintenance 



 

Warning 

For safe operation and maintenance of the product 

and to prevent possible electrical shock, fire, or 

personal injury: 
•  Repair the Product before use if the battery leaks. 
•  Be sure that the battery polarity is correct to 

prevent battery leakage. 

•  The battery contains hazardous chemicals that 

can cause burns or explode. If exposure to 

chemicals occurs, clean with water and get 

medical aid. 

•  Do not disassemble the battery. 
•  Do not disassemble or crush battery cells and 

battery packs. 

•  Do not put battery cells and battery packs near 

heat or fire. Do not put in sunlight. 

•  Do not short the battery terminals together. 
•  Do not keep cells or battery in a container where 

the terminals can be shorted. 

•  Remove the input signals before you clean the 

Product. 

•  Use only specified replacement parts. 
•  Have an approved technician repair the Product. 

Cleaning Outside Surfaces 

The DPM1B outside surfaces can be cleaned with a cloth dampened 
with water and mild detergent or alcohol. 

Battery Replacement 

The battery compartment of the Tester is found in the rear panel of the 
tester. Slide the compartment cover to expose the battery holder. 
Replace with one 9 V alkaline battery. 

 

Содержание DPM1B

Страница 1: ...ev 1 4 15 2006 2015 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies DPM1B Pneumatic Transducer...

Страница 2: ...SICAL DAMAGE CAUSED BY MISUSE OR PHYSICAL ABUSE IS NOT COVERED UNDER THE WARRANTY Items such as cables and nonserialized modules are not covered under this warranty Recalibration of instruments is not...

Страница 3: ...tches Technical Support For application support or answers to technical questions either email techservices flukebiomedical com or call 1 800 850 4608 or 1 440 248 9300 In Europe email techsupport eme...

Страница 4: ...recommends the product be calibrated at least once every 12 months Calibration must be done by qualified personnel Contact your local Fluke Biomedical representative for calibration Certification This...

Страница 5: ...s 5 Tester Familiarization 6 Front Panel 6 Back Panel 7 Accessories 7 Operation Overview 8 Operating the DPM1B 8 Setup 8 Zeroing the Monitor and Tester 9 Pressure Measurement 10 Testing Transducer Lin...

Страница 6: ...DPM1B Operators Manual ii...

Страница 7: ...iii List of Tables Table Title Page 1 Symbols 3 2 Accessories 7 3 Multiplication Factors 13...

Страница 8: ...DPM1B Operators Manual iv...

Страница 9: ...res Figure Title Page 1 DPM1B Front Panel 7 2 DPM1B to Monitor Setup 9 3 DPM1B Vented to Atmosphere 10 4 DPM1B Configured to Apply Pressure to Transducer 11 5 Sample Plot of Transducer Linearity Using...

Страница 10: ...DPM1B Operators Manual vi...

Страница 11: ...ctrical equipment Follow the manufacturer s electrical safety guidelines for equipment attached to the Tester A Warning identifies conditions and procedures that are dangerous to the user A Caution id...

Страница 12: ...Product Look for cracks or missing plastic Carefully look at the insulation around the terminals Use this Product indoors only Remove all probes test leads and accessories before the battery door is o...

Страница 13: ...ct Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information Manual Objectives Information about the design of the Tester to enable the user to underst...

Страница 14: ...surements necessary for determining the accuracy of any direct blood pressure monitoring system Vacuum negative and positive pressures from 300 mmHg to 300 mmHg can be generated within the Tester and...

Страница 15: ...casionally one or more of the elements involved in this process will malfunction causing inaccurate indications of blood pressure on the monitor Very often these errors can be attributed to a transduc...

Страница 16: ...ct USA FCC 47 CFR 15 subpart B Tester Familiarization The Tester incorporates a highly accurate digital pressure meter and pneumatic pressure generating cylinder into a small hand held package for det...

Страница 17: ...ure 1 DPM1B Front Panel Back Panel The back panel of the Tester houses the storage compartment for the 9 V alkaline battery Accessories Table 2 is a list of accessories available for the Tester Table...

Страница 18: ...ducer that is not performing as specified however could have other problems and sensitivity measurements can be an early indication of performance deterioration Setup 1 Connect the Tester transducer m...

Страница 19: ...pcock so that the two opposing end arrows point straight up and down and the middle arrow points toward the Tester as depicted in Figure 2 6 After filling the system position both Stopcocks as shown i...

Страница 20: ...asurement 1 Position the stopcock as shown in Figure 4 2 Move the thumbwheel of the Tester to change the static pressure generated See Figure 1 for location of thumbwheel The thumbwheel is rotated dow...

Страница 21: ...t of transducer linearity determined by using the Tester Pneumatic Transducer Tester 1 Test the transducer for linear operation in the range of 20 mmHg to 190 mmHg for example by taking measurements i...

Страница 22: ...eading 2 Record the pressure registered on the pressure monitor 3 Calculate the minimum and maximum pressure readings by using the transducer s sensitivity tolerance specifications For example For a 1...

Страница 23: ...er is not designed to generate the high pressures required by some of these instruments Multiplication Factors Table 3 Multiplication Factors PSI lb in 2 In Hg 0 C In H20 20 C mmHg 0 C mm H20 4 C ATM...

Страница 24: ...cells and battery packs Do not put battery cells and battery packs near heat or fire Do not put in sunlight Do not short the battery terminals together Do not keep cells or battery in a container wher...

Отзывы: