Operators Manual 

12 

Transducer Sensitivity Test 

Only one measurement is required to determine whether or not a 
transducer under test meets the manufacturer’s specified sensitivity. 
The Tester is used to generate a known pressure (generally 100 mmHg 
or higher works best). The excitation voltage of the transducer is 
determined from the manufacturer’s specifications. Transducer 
sensitivity is calculated as follows: 

1.  Apply 100 mmHg to the transducer as indicated by the Tester 

LCD reading. 

2.  Record the pressure registered on the pressure monitor 

3.  Calculate the minimum and maximum pressure readings by 

using the transducer’s sensitivity tolerance specifications. 
 
For example: For a 100 mmHg applied pressure with a 1 % 
tolerance, the minimum and maximum readings are 99 mmHg 
and 101 mmHg, respectively. 

4.  Compare the readings from Step 2, along with the results from 

Step 3. If the readings from Step 2 are within the range 
calculated in Step 3, the transducer is within the manufacturer’s 
specifications. 

Linear Plot

Nonlinear Plot

200

150

DPM-1B
Readings
(mmHg)

Monitor Readings(mmHg)

100

50

50

100

150

200

 

ehf5.eps 

Figure 5. Sample Plot of Transducer Linearity Using the DPM1B 

Содержание DPM1B

Страница 1: ...ev 1 4 15 2006 2015 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies DPM1B Pneumatic Transducer...

Страница 2: ...SICAL DAMAGE CAUSED BY MISUSE OR PHYSICAL ABUSE IS NOT COVERED UNDER THE WARRANTY Items such as cables and nonserialized modules are not covered under this warranty Recalibration of instruments is not...

Страница 3: ...tches Technical Support For application support or answers to technical questions either email techservices flukebiomedical com or call 1 800 850 4608 or 1 440 248 9300 In Europe email techsupport eme...

Страница 4: ...recommends the product be calibrated at least once every 12 months Calibration must be done by qualified personnel Contact your local Fluke Biomedical representative for calibration Certification This...

Страница 5: ...s 5 Tester Familiarization 6 Front Panel 6 Back Panel 7 Accessories 7 Operation Overview 8 Operating the DPM1B 8 Setup 8 Zeroing the Monitor and Tester 9 Pressure Measurement 10 Testing Transducer Lin...

Страница 6: ...DPM1B Operators Manual ii...

Страница 7: ...iii List of Tables Table Title Page 1 Symbols 3 2 Accessories 7 3 Multiplication Factors 13...

Страница 8: ...DPM1B Operators Manual iv...

Страница 9: ...res Figure Title Page 1 DPM1B Front Panel 7 2 DPM1B to Monitor Setup 9 3 DPM1B Vented to Atmosphere 10 4 DPM1B Configured to Apply Pressure to Transducer 11 5 Sample Plot of Transducer Linearity Using...

Страница 10: ...DPM1B Operators Manual vi...

Страница 11: ...ctrical equipment Follow the manufacturer s electrical safety guidelines for equipment attached to the Tester A Warning identifies conditions and procedures that are dangerous to the user A Caution id...

Страница 12: ...Product Look for cracks or missing plastic Carefully look at the insulation around the terminals Use this Product indoors only Remove all probes test leads and accessories before the battery door is o...

Страница 13: ...ct Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information Manual Objectives Information about the design of the Tester to enable the user to underst...

Страница 14: ...surements necessary for determining the accuracy of any direct blood pressure monitoring system Vacuum negative and positive pressures from 300 mmHg to 300 mmHg can be generated within the Tester and...

Страница 15: ...casionally one or more of the elements involved in this process will malfunction causing inaccurate indications of blood pressure on the monitor Very often these errors can be attributed to a transduc...

Страница 16: ...ct USA FCC 47 CFR 15 subpart B Tester Familiarization The Tester incorporates a highly accurate digital pressure meter and pneumatic pressure generating cylinder into a small hand held package for det...

Страница 17: ...ure 1 DPM1B Front Panel Back Panel The back panel of the Tester houses the storage compartment for the 9 V alkaline battery Accessories Table 2 is a list of accessories available for the Tester Table...

Страница 18: ...ducer that is not performing as specified however could have other problems and sensitivity measurements can be an early indication of performance deterioration Setup 1 Connect the Tester transducer m...

Страница 19: ...pcock so that the two opposing end arrows point straight up and down and the middle arrow points toward the Tester as depicted in Figure 2 6 After filling the system position both Stopcocks as shown i...

Страница 20: ...asurement 1 Position the stopcock as shown in Figure 4 2 Move the thumbwheel of the Tester to change the static pressure generated See Figure 1 for location of thumbwheel The thumbwheel is rotated dow...

Страница 21: ...t of transducer linearity determined by using the Tester Pneumatic Transducer Tester 1 Test the transducer for linear operation in the range of 20 mmHg to 190 mmHg for example by taking measurements i...

Страница 22: ...eading 2 Record the pressure registered on the pressure monitor 3 Calculate the minimum and maximum pressure readings by using the transducer s sensitivity tolerance specifications For example For a 1...

Страница 23: ...er is not designed to generate the high pressures required by some of these instruments Multiplication Factors Table 3 Multiplication Factors PSI lb in 2 In Hg 0 C In H20 20 C mmHg 0 C mm H20 4 C ATM...

Страница 24: ...cells and battery packs Do not put battery cells and battery packs near heat or fire Do not put in sunlight Do not short the battery terminals together Do not keep cells or battery in a container wher...

Отзывы: