LI5X Process & Temperature Meter                       Instruction Manual 

61

 

 

Tips: 

• 

Low and high input signals can be any valid values within the 
range of the meter. 

• 

Observe minimum input span requirements between input 1 
and input 2. 

• 

Low input must be less than high input signal. 

Error Message (Err) 

An error message indicates that the calibration or scaling process was 
not successful.  
After the error message is displayed, the meter reverts to input 1, allow-
ing the appropriate input signals to be applied. 
 
The error message might be caused by any of the following conditions: 
1.  Input signal is not connected to the proper terminals, or it is con-

nected backwards. 

2.  Wrong signal selection in 

Setup

 menu. 

3.  Minimum input span requirements not maintained. 
 

Minimum Input Span 

The minimum input span is the minimum difference between input 1 and 
input 2 signals required to complete the calibration or scaling of the meter. 
 

Input range 

Input 1 & input 2 span 

4-20 mA 

0.40 mA 

±

10 VDC 

0.20 VDC 

 

Spec Tech Industrial Electric

               www.spectechind.com

888-773-2832

Spec Tech Industrial Electric

               www.spectechind.com

888-773-2832

Содержание LI50-1001

Страница 1: ...play UL Type 4X NEMA 4X IP65 Front Universal Power Supply 85 265 VAC 12 36 VDC 12 24 VAC Power Option Two Relays or 4 20 mA Output Option 24 VDC Transmitter Power Supply Options Serial Communication A...

Страница 2: ...ll trademarks mentioned in this document are the property of their re spective owners Visit our Web Site http www flowline com Spec Tech Industrial Electric www spectechind com 888 773 2832 Spec Tech...

Страница 3: ...excellent addition to any system ORDERING INFORMATION Model Options LI50 1001 24 VDC Transmitter Power Supply LI50 1001 MB 24 VDC Transmitter Power Supply Modbus LI51 1001 2 Relays 24 VDC Transmitter...

Страница 4: ...15 Power Connections 15 Signal Connections 16 Serial Communication 19 Relays and 24 V Output Connections 19 Switching Inductive Loads 20 4 20 mA Output Input Signal Connections 21 SETUP AND PROGRAMMI...

Страница 5: ...es 51 Offset Adjustment Adj 53 Noise Filter fltr 53 Noise Filter Bypass byps 54 Serial Communications serl 54 Select Menu SElc 55 Linear or Square Root Function linr or Sqrt 55 Low Flow Cutoff cutF 56...

Страница 6: ...ure 8 Three Wire RTD Input Connections 18 Figure 9 Two Wire RTD Input Connections 18 Figure 10 Four Wire RTD Input Connections 19 Figure 11 Relay 24 V Output Connections 19 Figure 12 AC and DC Loads P...

Страница 7: ...gs reached by the process are stored until reset by the user or until power to the meter is turned off PASSWORD Programmable password restricts modification of programmed settings NON VOLATILE MEMORY...

Страница 8: ...ears parts and labor Process Input INPUTS Field selectable 20 mADC 0 20 4 20 mA and 10 VDC 0 5 1 5 0 10 V ACCURACY 0 05 of span 1 count square root 10 100 F S LOW FLOW CUTOFF 0 9999 0 disables cutoff...

Страница 9: ...to 1578 F 50 to 870 C 2 F 1 C 100 RTD 328 to 1382 F 200 to 750 C 1 F 1 C COLD JUNCTION REFERENCE Automatic fixed no user calibration needed TEMPERATURE DRIFT 2 C maximum 0 to 65 C ambient temperature...

Страница 10: ...tic reset only non latching Automatic manual reset at any time non latching Manual reset only at any time latching Manual reset only after alarm condition has cleared latching Automatic reset Relays w...

Страница 11: ...CE Power supply Minimum Maximum 24 VDC 10 700 35 VDC external 100 1200 Serial Communications METER ADDRESS Standard Proprietary Data Comm PDC protocol 0 99 Optional Modbus protocol 1 247 BAUD RATE 300...

Страница 12: ...adiated Emissions Class A AC Mains Conducted Emissions Class A IMMUNITY EN 61000 6 2 2001 EMC heavy industrial generic immunity standard RFI Amplitude Modulated 80 1000 MHz 10 V m 80 AM 1 kHz Electric...

Страница 13: ...mize EMC performance Declaration of Conformity available SAFETY INFORMATION CAUTION Read complete instructions prior to instal lation and operation of the meter WARNING Risk of electric shock WARNING...

Страница 14: ...0 06 1 5 mm steel panel 0 16 4 1 mm plastic panel Remove the two mounting brackets provided with the meter back off the two screws so that there is 6 4 mm or less through the bracket Slide the bracket...

Страница 15: ...lable with requested configuration It also identifies the location of the RTD TC selector switch 24V OUT EL1601 POWER EXC T V mA COM RT D S WIT CH T C T C RT D 1 2 3 4 5 6 COM NO NO NC NC COM RELAY2 R...

Страница 16: ...3 5 4 V T EXC 2 Wire 4 20 mA Self Powered Transmitter Figure 4 Transmitter Powered by Ext Supply or Self Powered COM SIGNAL CONNECTOR mA 2 1 3 5 4 V T EXC 2 Wire 4 20 mA Transmitter P P 2 1 24V OUT C...

Страница 17: ...ures show examples for thermocouple and RTD con nections The RTD TC selector switch must be set to the proper position for the meter to accept the selected temperature input The input type is selected...

Страница 18: ...he three wire RTD connection has built in lead wire compensation SIGNAL CONNECTOR 2 1 3 5 4 RTD TC 1 2 3 4 5 6 COM mA V T EXC Switch Position RTD Sensor Figure 9 Two Wire RTD Input Connections Lead wi...

Страница 19: ...2 Use LI97 1001 232 for RS 232 interfacing Use a standard modular cable for meter to meter interfacing for cloning purposes i e copying programmed settings from one meter to other meters Relays and 24...

Страница 20: ...Choose R and C as follows R 0 5 to 1 for each volt across the contacts C 0 5 to 1 F for each amp through closed contacts Notes 1 Use capacitors rated for 250 VAC 2 RC networks may affect load release...

Страница 21: ...om an internal power supply optional or from an external power supply I mA OUT P1 4 5 3 P1 24V OUT 6 I Remote Display Chart Recorder etc Figure 14 4 20 mA Output Input Signal Powered by Meter External...

Страница 22: ...s involved in the setup process of the meter The RTD TC selector switch located between the SIGNAL and SERIAL connectors must be set accordingly for the meter to accept RTD or thermocouple inputs Figu...

Страница 23: ...ny time Press the Right arrow button to move to the next digit during digit programming Press the Up arrow button to scroll through the menus decimal point or to increment the value of a digit Press t...

Страница 24: ...e 100 RTD A392 Alpha 392 Set 0 00392 American curve 100 RTD tC TC Set meter for TC input 0 J 0 J Type J 1 k 1 K Type K 2 T 2 T Type T 3 t 0 3 T 0 Type T 0 1 resolution 4 E 4 E Type E f C F or C Set te...

Страница 25: ...1 fail safe operation on On Enable fail safe operation FLS2 Fail safe2 Set relay 2 fail safe operation DLAY Delay Enter Time Delay menu DLY1 Delay1 Enter relay 1 time delay setup On1 On1 Set relay 1...

Страница 26: ...alue dis2 Display 2 Program display 2 value err Error Error calibration not successful check signal pass Password Enter the Password menu unlC Unlocked Program password to lock meter loCd Locked Enter...

Страница 27: ...9846 setu prog pass Run Mode Press Menu at any time to exit and return to Run Mode Changes made to settings prior to pressing Enter Ack are not saved Changes to the settings are saved to memory only a...

Страница 28: ...ed is displayed brighter than the rest Press the Enter Ack button at any time to accept a setting or Menu button to exit without saving changes 04 00 04 00 05 00 Increment Digit Value Next Setting Sel...

Страница 29: ...p arrow but ton to scroll through choices Press the Menu button to exit at any time setu inpt dec p F C rely Aout Relay menu is always available even if relays option is not installed Visual alarm ind...

Страница 30: ...ake Selections Press Up Arrow to Scroll Through Choices Press Menu to Exit at any Time J K T T 0 E If RTD is selected the display shows A385 or A392 Select the coeffi cient to match the RTD sensor eit...

Страница 31: ...mal point one place to the right until no decimal point is displayed then it moves to the leftmost position dec p dddd Next Select Decimal Point Previous Setting the Temperature Scale F C The meter ca...

Страница 32: ...l automatic reset only f Off relay and status LED disabled 2 Set point 3 Reset point 4 Fail safe operation a On enabled b Off disabled 5 Time delay a On delay 0 199 seconds b Off delay 0 199 seconds r...

Страница 33: ...p in a similar fashion Auto A m LtCH Act1 L CL From Relay 1 Menu Press Enter Ack button to accept setting Press Menu button to exit at any time ALtr oFF Programming Set and Reset Points High alarm ind...

Страница 34: ...Programming Time Delay The On and Off time delays may be programmed for each relay be tween 0 and 199 seconds The relays will transfer only after the condi tion has been maintained for the correspondi...

Страница 35: ...ff on Relay LED Automatic or Manual non latching ACK pressed Relay LED Manual latching ACK pressed Relay LED Manual only after passing below Reset latching with clear ACK pressed Set Reset For Manual...

Страница 36: ...atching ACK pressed Relay LED Manual latching ACK pressed Relay LED Manual only after passing above Reset latching ACK pressed For Manual reset mode ACK can be pressed anytime to turn off relay For re...

Страница 37: ...starts and the relay trips when the time delay has elapsed If the signal drops below the set point high alarm before the time delay has elapsed the On time delay timer resets and the relay does not ch...

Страница 38: ...atic or Manual non latching ACK pressed Relay LED Manual latching ACK pressed Relay LED Manual only after passing above Reset latching ACK pressed Fail safe operation relay coil is energized in non al...

Страница 39: ...r Manual non latching ACK pressed Relay LED Manual latching ACK pressed Relay LED Manual only after passing above Reset latching with clear ACK pressed Fail safe operation relay coil is energized in n...

Страница 40: ...t de energized energized Relay1 Set Relay2 Set Relay1 Relay2 Reset LEDs indicate the relay status Input Input Relay1 Reset Relay2 Reset Relay1 Reset Pump 2 Pump 1 Pump 2 Pump 1 Pump 2 de energized ene...

Страница 41: ...Output menu is used to program 1 4 20 mA output based on display values 2 Sensor break value in mA out1 04 00 Press Enter to Accept Setting dis2 Scale Output 2 dis1 Set Display 1 Value scal 00 00 Set...

Страница 42: ...00 to 23 00 mA e g If sensor break value is programmed to 0 00 the actual output will not be greater than 1 00 mA sEbr Set Sensor Break Value scal 03 00 Aout Program Menu Press Enter to Accept Setting...

Страница 43: ...e g 4 20 mA A cali brated signal source is not needed to scale the meter For thermocouple and RTDs just connect the sensor to the proper terminals and turn the power on No calibration needed when the...

Страница 44: ...s dis1 00 0 Press Enter to Accept Setting inp2 Scale Input 2 inp1 Set Input 1 Value scal 04 00 Set Display 1 Value Press Up to Set Digit Value Press Right to Select Next Digit Press Menu to Exit at an...

Страница 45: ...he proper terminals or it is con nected backwards 2 Wrong signal selection in Setup menu 3 Minimum input span requirements not maintained 4 Input 1 signal inadvertently applied to calibrate input 2 Mi...

Страница 46: ...00 Set Display 1 Value inp2 Calibrate Input 2 Cal Display Flashes Accepting Input See Scaling the Meter for Button Functions Description Recalibrating Temperature Inputs Cal Remember the meter is cali...

Страница 47: ...appropriate wire e g type J thermocouple wire to recalibrate type J input see page 17 2 Set up the meter to accept the selected input e g type J T C see page 30 3 Set up the meter to display temperat...

Страница 48: ...s see Setting Numeric Values page 28 pass unlC 0000 Program Password loCd Run Mode Record the password for future reference If appropriate it may be re corded in the space provided Model Serial Number...

Страница 49: ...conds then it returns to Run Mode To try again press Enter Ack while the Locked message is displayed Forgot the Password The password may be disabled by the following procedure 1 Note display reading...

Страница 50: ...res of the meter Press and hold Right arrow first then Menu Copy I Cal Press Enter Ack to Access Menu or to Accept Setting diag byps fltr serl Press Up to Scroll Menu and to Increment Digit Value Pres...

Страница 51: ...ram byte to byte timeout silent time Modbus only Copy Copy Enter copy function send Send Send meter settings to another meter done Done Copy function completed Selc Select Enter the Select menu functi...

Страница 52: ...ettings led LED Test display CJC CJC Display cold junction compensation voltage Cfg CFG Display meter configuration pts Points Display calibration points for process inputs RELY Relays Display relay s...

Страница 53: ...matic see note 2 below for important limitations Notes 1 Offset adjustment is available only when TC or RTD input is se lected 2 If adjustment value is greater than 11 C and the temperature scale is...

Страница 54: ...g Flowline Proprietary Data Communication PDC protocol stan dard The Modbus RTU protocol is optional To communicate with a computer or other data terminal equipment an RS 232 or RS 422 485 adapter opt...

Страница 55: ...put selection menu is for factory use only Do not change output selection inty SelectMenu Linear or Square Root Function linr or Sqrt Meters are set up at the factory for linear function The linear fu...

Страница 56: ...cutoff cutF Set Cutoff Value Func 03 00 SELc Press Enter to Accept Setting Press Up to Set Digit Value Press Right to Select Next Digit Press Menu to Exit at any Time Advanced Features out Display In...

Страница 57: ...py The Copy function is used to copy or clone all the settings from one meter to other meters requiring exactly the same setup and program ming i e type of input scaling decimal point filter bypass et...

Страница 58: ...hen press Enter Ack 5 The meter displays the message Send Press Enter Ack the dis play flashes while sending data The message done is displayed when copying is completed Copy send Advanced Features do...

Страница 59: ...parately if meter will be used with all input types Notes If meter is in operation and it is intended to accept only one input type e g 4 20 mA recalibration of other inputs is not necessary Allow the...

Страница 60: ...display value to correspond to the input signal being cali brated 7 The display moves to the high input calibration I Hi Apply the high input signal and press Enter Ack 8 Set the display for the high...

Страница 61: ...ppropriate input signals to be applied The error message might be caused by any of the following conditions 1 Input signal is not connected to the proper terminals or it is con nected backwards 2 Wron...

Страница 62: ...nd temperature units F or C selected Type T thermocouples can be displayed with either 1 or 0 1 resolution Front Panel Buttons Operation Button Symbol Description Press to enter or exit Programming Mo...

Страница 63: ...splay maximum reading since the last reset power up 2 Press Up arrow Max again to display the minimum reading since the last reset power up 3 Press Enter Ack to hold Max Min display reading the meter...

Страница 64: ...76 45mm 0 59 15mm 3 2 81mm 2 45 62mm 3 6 91mm Figure 17 Meter Dimensions Side View 3 61 92mm 2 50 64mm 4 68 119mm Figure 18 Case Dimensions Top View Spec Tech Industrial Electric www spectechind com 8...

Страница 65: ...eshooting purposes Press the Enter Ack button to view the set tings and the Menu button to exit at any time For a description of the diagnostics messages see Advanced Features Menu Display Messages pa...

Страница 66: ...Press Up arrow to go to Diagnostics menu 3 Press and hold Right arrow Reset for five seconds press Enter Ack when display flashes reset Note If Enter Ack is not pressed within three seconds display r...

Страница 67: ..._________ S N _______________ Date _________ Parameter Display Default Setting User Setting Input type inpt 4 20 mA Programming prog Scale Input 1 InP1 4 00 mA Display 1 Dis1 4 00 Input 2 InP2 20 00 m...

Страница 68: ...value CutF 0 00 disabled Output option out Factory set only Display intensity inty Level 2 Modbus defaults Address addr 247 Parity prty Even Byte to byte timeout tbyt 0 01 sec Note The byte to byte t...

Страница 69: ...adings Display response is too slow Check filter and bypass values Inaccurate temperature reading Check 1 Temperature units F or C 2 TC type or RTD curve selected 3 Offset adjustment 4 TC wire used 5...

Страница 70: ...F or C To Program Locked Unlocked Calibrate Scale To Password TC or RTD Yes No TC or RTD Yes No Relays Relay 1 Relay 2 Failsafe Delays Action Set point Reset point Auto Auto Manual Latch Latch w Clea...

Страница 71: ...e Pressing Enter Ack will disable the 10 second timeout and continuously display Max or Min Press hold Right arrow and Menu for 3 seconds to access Advanced Features Menu When the meter is locked Adju...

Страница 72: ...Call 562 598 3015 Fax 562 431 8507 Email US sales flowline com Web Site www flowline com For International Sales and Customer Service Call 562 598 3015 Fax 562 431 8507 Email Int sales flowline com W...

Отзывы: