EchoPod
31 MAR 11
MN204210
Rev B
Factory Defaults Table
Testing the Transmitter
1.
Connect a multimeter in series with the black wire to read the current output.
2.
Verify that the current increases (tank filling) and decreases (tank emptying) appropriately
in the calibrated span.
3.
If not, carefully observe and attempt to correlate any installation, level or application event
for more specific troubleshooting direction.