22 of 29
EchoPod
31 MAR 11
MN204210
Rev B
Tank Adapter
w/ 2”x1” Reducer Bushing
Tank Adapter
Thread x Thread
Do not use thread x thread
2.
Riser:
Installations with tall, narrow risers can impede the acoustic signal.
a.
DL34 Series:
2” (5 cm) diameter risers should be no taller than 5” (12.7 cm).
Larger diameter risers should be no taller than 12” (30.5 cm).
b.
DL14 & DL24 Series:
Note:
Do not exceed the dimensions listed above
3.
Flange (DL14 & DL24 series):
If installing on a flange, select a flange with a thread that is above the plane of the flange,
such as the LM52-1850.
a.
The DL34 series works well with Flange installations.
b.
Avoid the use of blind flanges with tapped threads or flanges where the threads are
even with the plane of the flange, such as the Banjo 1" Poly ANSI Flange (series
AF100).
c.
Use a flange with a 2” thread and add a 2” to 1” reducer bushing to complete the
installation.
1” Flange w/
thread in plane
2” Flange w/
reducer bushing
Do not use thread in plane