19 Instruction Manual
M21
Window for entering the fluid sonic velocity only for non-standard liquids
M22
Window for entering the viscosity of the non-standard liquids
M23
Window for selecting the proper transducers
There are 14 different types of transducers for selection.
If the user-type-transducers are used, 4 user type wedge parameters,
which will be prompted by the software, should be entered following.
If the
π
type transducers are used, 3
π
type transducers and pipe
parameters should be entered following.
M24
Window for selecting the transducer mounting methods
Four methods can be selected:
(0) V-method (1) Z-method (2) N-method (3) W-method
M25
Display the transducer mounting spacing
M26
Entry to store the parameter configuration into the internal NVRAM
M27
Entry to load one set of saved parameters
M28
Select YES or NO for the instrument to determine whether or not to hold
(or to keep) the last correct value when poor signal condition occurs. YES
is the default setup
M29
Enter a value ranging from 000 to 999. 0 is the default value
M30
Window for selecting unit system. Default value is ‘Metric’. The change
from English to Metric or vice versa will not affect the unit for totalizers.
M31
Window for selecting flow rate that will be used by the instrument
afterward.
Flow rate can be in
0. Cubic meter short for (m3)
1. Liter (l)
2. USA gallon (gal)
3. Imperial Gallon (igl)
4. Million USA gallon (mgl)
5. Cubic feet (cf)
6. USA liquid barrel (bal)
7. Imperial liquid barrel (ib)
8. Oil barrel (ob)
The flow unit in terms of time can be per day, per hour, per minute or per
second. So there are 36 different flow rate units in total for selection.
M32
Window for selecting the totaliziers’ working unit
M33
Select totalizer multiplier
The multiplier ranges from 0.001 to 10000
M34
Turn on or turn off the NET totalizer
M35
Turn on or turn off the POS totalizer
M36
Turn on or turn off the NEG totalizer
M37
(1) Totalizer reset