1.
Product Instruction
M6 series pump speed range is 0.1-600 rpm, flow rate range is 0.211~12000 ml/min.
Adjust speed manually or automatically through external control interface. Power-down memory function, when re-power on, the pump will
work with previous status. Timing test function, timing range: 0.5~999s, it is convenient for calibration and test. It can support 7 different
external control modes, make more convenient.
2.
Product Appearance
A- Drive
B- Pump head
C- External control interface
D- Power socket
E- Internal fuse
F- Power switch
3.
Product Keypad Instruction
l
Start /stop button
Control the pump start and stop, press this button once the pump work status change once. When the pump runs, the start/stop indicator light
is on. When the pump stops, the start/stop indicator light is off.
l
Direction button
Control the pump working direction, press the button once, the drive work status change once. When the pump runs left, the left indicator
light is on. When the pump runs right, the right indicator light is on.
l
Digital knob
(1)
Turn it left or right is to adjust the motor speed when displays the speed.
(2)
Press down the knob then release in stop status, enter RS485 address setting interface, then a number occur and flashes, now turn the
knob to change address, press the knob again to confirm and escape from this setting.