4.3.1 Basic Configuration Interface
Operating steps:
a.
Press
[Knob]
to select tubing in setting interface, when in status (1)
, rotate
[Knob]
, the cursor up and down, select the tubing or
back suction angle.
b.
When in status (1), press
[Knob]
button, change to status (2).
c.
When in status (2), rotate
[Knob]
button, select tubing.
d.
When in status (2), press
[Knob]
back to status (1), ensure the inputting value.
e.
When in status (1), cursor moves to Back, press
[Knob]
to back to menu selection interface.
Set back suction angel, the angle range is 0-360° (360° is a full circle of back suction). When the angle is 0, it means turning off back
suction function.
Note
: When turn off back suction angle, the main interface does not display the mark of back suction angle. Only when turn on it, the
main interface will display the mark.
4.3.2 Flow Calibration Interface
Operation steps:
a.
After enter the interface, the default test time is 60 seconds, the actual liquid volume defaults to the liquid volume displayed
according to the current set flow.
b.
Press
[Knob]
button to select test time, when in status (1), rotate
[Knob]
and the cursor moves to actual volume line, then press
[Start/stop]
button, the motor start running, and start calibration. The countdown is displayed at the test time.
c.
When in status (1), press
[Knob]
button, switch to status (2), when in this status, rotate
[Knob]
button, set the test time.
d.
After the countdown, the cursor move to actual volume directly, input the actual volume with rotating knob.
e.
Also can directly choose actual volume box, press
[Knob]
button to set the actual liquid volume.
f.
After input, press
[Knob]
, ensure inputting, a dialog box with “Calibration complete” will pop up in this interface, that means
completing calibration. Press
[Knob]
button back to calibration interface.
g.
Reset the calibration function: Choose
[Yes]
, press
[Knob]
button, a dialog box with “Reset calibration completes” will pop up in
this interface, that means the calibration data can be reset.
h.
After calibrating, then back to main interface, the speed will not change, the flow will change according to calibrating actual
volume. If the speed is 100rpm, the previous corresponding flow is 20mL/min, the actual liquid volume tested in 60 seconds is
30ml/min, after calibrating, the speed displayed in main interface is 100 rpm, the flow is 30L/min.
FD15
19#
1
360
Head&Tube
Head:
Tube: 13#
Suck back: 0
esc
KT15
FD15
19#
2
360
Head&Tube
Head:
Tube:
Suck back: 360
esc
KT15
19#
60.0
215.38 mL
1
60.0
215.38 mL
2
Calibration
Calibration
Time:
Sec
Volume:
Reset:
No
esc
Time:
Volume:
Reset:
esc
No
Sec