Flow Controller Systems FCS KA1 Скачать руководство пользователя страница 4

 

           

 

   

 

    KA1

- 1 - 

1.

 

Product Introduction 

KA1 adopts ABS engineering plastic housing, 2.4 inch LCD screen, low power, ultra-quiet, compact structure. Easy to operate with 

digital  knob  speed  control. A  variety  of  external  control  methods  for  option.  Support  RS485  communication,  standard  MODBUS 

communication protocol (RTU mode); It is suitable for complex working environment with super anti-interference and wide voltage 

design. 

Suitable pump head: KT15 pump head. 

2.

 

Product Appearance and Interface Instruction 

 

A——Drive 

B——Pump Head 

C——Fan 

D——External Control Interface 

E——USB Port 

F——Power Supply Interface 

 

 

3.

 

Operation Panel Instruction 

KA1 can meet variety of working environment with pure imported keyboard, which has 300000 times lives. 

 

l

 

Start/Stop Button

: Press 

[

Start/stop

]

 button, pumps run according to current speed, press 

[

Start/stop

]

 button again, pumps stop 

running. 

l

 

Reversing Button

:

 

Press 

[

reversing

]

 button once, pump will change running direction once. 

l

 

Full Speed Button

: When in stopping state or transferring state, press 

[

Full speed

]

, the pump will run with highest speed. This 

button can be used for washing tubing or fast filling liquid. 

l

 

Knob

: Rotate 

[

knob

]

 in main interface to adjust speed and flow will change. 

In stopping status, press 

[

knob

enter to 

[

Menu selection interface

]

, rotate 

[

knob

to select corresponding items. Press 

[

knob

to select parameter which needed to change when setting parameters. 

 

 

 

Knob

Start/stop

Full speed

Reversing

Содержание FCS KA1

Страница 1: ...FCS KA1 Compact Peristaltic Pump Operation Manual 0 1 2 3456 1 7 7 8 9 88 1 6 A B6 1...

Страница 2: ...f the power cord or plug had wear and other damage please disconnect the plug Hold the plug instead of the wire If following situations happened please turn off the power supply and disconnect the plu...

Страница 3: ...and Interface Instruction 1 3 Operation Panel Instruction 1 4 Operation Interface Structure 2 5 External Control Interface 7 6 Technical Parameters 8 7 Main Function and Feature 9 8 Dimension Drawing...

Страница 4: ...tion KA1 can meet variety of working environment with pure imported keyboard which has 300000 times lives l Start Stop Button Press Start stop button pumps run according to current speed press Start s...

Страница 5: ...n stop status Direction The screen displays when in clockwise displays in counterclockwise Flow speed displays Displays the flow value with the current pump head tubing and speed Parameters Displays a...

Страница 6: ...tart running and start calibration The countdown is displayed at the test time c When in status 1 press Knob button switch to status 2 when in this status rotate Knob button set the test time d After...

Страница 7: ...us 1 and ensure items The product support Modbus communication protocol RTU mode communication interface RS485 Baud rate means Symbol Baud rate Data bits Stop bit Check bit 9600 8 e 1 9600 8 1 Parity...

Страница 8: ...s 1 rotate Knob button the cursor moves up and down to select external control speed signal maximum speed b When in status 1 press Knob button to status 2 c When in status 2 rotate Knob button to sele...

Страница 9: ...signal When the motor speed and analog signal reach the set maximum speed according to the corresponding proportional relationship the analog signal will be increased at this time and the motor speed...

Страница 10: ...p port In pulse mode pin1 is short circuited with pin14 and then disconnect the motor will run do it one more time the motor will stop In level mode pin1 is short circuited with pin14 the motor will r...

Страница 11: ...3 Start motor the relay pull in the motor stop the relay disconnect l The square Port USB interface is shown in the figure below Square Port USB interface equipped with square port data cable can be u...

Страница 12: ...lugs must be connected or disconnected in the power off status to prevent the external control interface from being burned The user s power socket must have ground wire and have reliable grounding Thi...

Страница 13: ...and pumping accessories as consumable items are excluded Electrical surge as a cause of failure is excluded Chemical attack is excluded Improper operation or man made damage as a cause of failure is e...

Отзывы: