- 7 - 

 

 

5.

 

External Control Interface 

The external control interface of 15 hole base is shown in the figure below: 

 

l

 

The external control interface is 15 hole base, the function of 15 hole pin is: 

Pin1 is the external control start/stop port; 

Pin2 is the external control reversing port; 

Pin3 is empty. 

Pin4 is RS485 B-; 

Pin5 is ; 

Pin6 is 0-5V signal terminal of analog speed regulation; 

Pin7 is 0-10V signal terminal of analog speed regulation; 

Pin8 is 4-20mA signal terminal of analog speed regulation; 

Pin9 is AGND common port of analog speed regulation; 

Pin10 is empty; 

Pin11 is negative terminal (POWER_GND) of status output; 

Pin12 is status output terminal; 

Pin13 is positive terminal (POWER_VDD) of status output; 

Pin14 is the positive terminal of internal isolated 5V output; 

Pin15 is the negative terminal of internal isolated 5V output. 

(1)

 

External control start/stop, reversing: 

Set external control mode in external control setting interface, turn on corresponding external control function, the external control signal 

input port is valid. 

Ø

 

Pin1 is the external control start/stop port

In pulse mode

, pin1 is short-circuited with pin14 and then disconnect, the motor will run, do it one more time, the motor will stop; 

In 

level mode

, pin1 is short-circuited with pin14, the motor will run, disconnect them, the motor will stop. 

Ø

 

Pin2 is the external control reversing port: 

In pulse mode

, pin2 is short-circuited with pin14 and then disconnect them once, the motor direction will change once; 

In level mode

pin2 is short-circuited with pin14, the motor will run with clockwise, disconnect them, the motor will run with counterclockwise. 

Ø

 

Pin14 is the common port of external control start/stop, reversing 

Note:   

1)

 

When the external control is turned on level mode, the membrane keypad will not make effect. 

2)

 

The pump defaults to passive external control signal, the above description is also the default state. If you need active 

external control signal, please contact with manufacture. 

(2)

 

RS485 communication port: 

Select pump address and communication baud rate in communication setting interface.

 

Ø

 

Pin5 RS485A:

 Connect RS485 A+ port

 

Ø

 

Pin4 RS485B:

 Connect RS485 B- port

 

Note: The communication protocol is standard MODBUS protocol. 

(3)

 

Analog signal input port:

 After selecting external control speed regulation signal in external control setting interface, turn on the 

1

15

5

6

10

11

Содержание FCS KA1

Страница 1: ...FCS KA1 Compact Peristaltic Pump Operation Manual 0 1 2 3456 1 7 7 8 9 88 1 6 A B6 1...

Страница 2: ...f the power cord or plug had wear and other damage please disconnect the plug Hold the plug instead of the wire If following situations happened please turn off the power supply and disconnect the plu...

Страница 3: ...and Interface Instruction 1 3 Operation Panel Instruction 1 4 Operation Interface Structure 2 5 External Control Interface 7 6 Technical Parameters 8 7 Main Function and Feature 9 8 Dimension Drawing...

Страница 4: ...tion KA1 can meet variety of working environment with pure imported keyboard which has 300000 times lives l Start Stop Button Press Start stop button pumps run according to current speed press Start s...

Страница 5: ...n stop status Direction The screen displays when in clockwise displays in counterclockwise Flow speed displays Displays the flow value with the current pump head tubing and speed Parameters Displays a...

Страница 6: ...tart running and start calibration The countdown is displayed at the test time c When in status 1 press Knob button switch to status 2 when in this status rotate Knob button set the test time d After...

Страница 7: ...us 1 and ensure items The product support Modbus communication protocol RTU mode communication interface RS485 Baud rate means Symbol Baud rate Data bits Stop bit Check bit 9600 8 e 1 9600 8 1 Parity...

Страница 8: ...s 1 rotate Knob button the cursor moves up and down to select external control speed signal maximum speed b When in status 1 press Knob button to status 2 c When in status 2 rotate Knob button to sele...

Страница 9: ...signal When the motor speed and analog signal reach the set maximum speed according to the corresponding proportional relationship the analog signal will be increased at this time and the motor speed...

Страница 10: ...p port In pulse mode pin1 is short circuited with pin14 and then disconnect the motor will run do it one more time the motor will stop In level mode pin1 is short circuited with pin14 the motor will r...

Страница 11: ...3 Start motor the relay pull in the motor stop the relay disconnect l The square Port USB interface is shown in the figure below Square Port USB interface equipped with square port data cable can be u...

Страница 12: ...lugs must be connected or disconnected in the power off status to prevent the external control interface from being burned The user s power socket must have ground wire and have reliable grounding Thi...

Страница 13: ...and pumping accessories as consumable items are excluded Electrical surge as a cause of failure is excluded Chemical attack is excluded Improper operation or man made damage as a cause of failure is e...

Отзывы: