BE6000 Ultrasonic Flow Meter O&M Manual
Flomotion Systems, Inc. ~ FlomotionSystems.com
36
Therefore, 00 indicates the poorest signal while 99 indicates the best signal. Normally signal quality Q value
should be better than 60.
•
TOM/TOS*100 (M91)
Display the ratio between the actual measured transmit time and the calculated transmit time according to
customer’s requirement. Normally the ratio should be 100 ±3%. If the difference is too large, the user should
check if the parameters are entered correctly, especially the sound velocity of the fluid and the installation of
the transducers.
This data is of no use before the system is ready.
•
Fluid Sound Velocity (M92)
Display the measured fluid sound velocity. Normally this value should be approximately equal to the entered
value in M21. If the difference is too large, it probably results from an incorrect value entered in M21 or
improper installation of the transducers.
•
Total Time and Delta Time (M93)
Display the measured ultrasonic average time (unit:
μ
S) and delta time of the upstream and downstream
(unit: nS) time. The velocity calculation in BE6000 ultrasonic flowmeter is based on the two readings;
especially the delta time will best indicate if the instrument is running steadily.
Normally the fluctuation in the ration of the delta time should be lower than 20%; otherwise, the system may
not run steadily. It is, then, necessary to check if the transducers are installed properly or if the parameters
have been entered correctly.
•
Reynolds Number and Factor (M94)
Displays the Reynolds number that is calculated by the system and the factor that is set currently by the
system. Normally this scaling factor is the average of line and surface velocity factor inside the pipe.
•
Print Commands
(M97, M98, M99, M9.)
Press [MENU] [9] [7]. Prints the following working parameters as set by the user at the initial set up;
-
Outer Diameter
-
Wall Thickness
Reynolds Number [94
0.0000 1.0000
Total Time, Delta Time
8.9149uS, -171.09nS
Fluid Sound Velocity
0.0000 m/s
TOM/TOS*100 [91
0.0000