MN16 V1.2 7/08
13
Function Range Resolution
Accuracy
5.000Hz 0.001Hz
50.00Hz 0.01Hz
±
(1.5% r 5 digits)
500.0Hz 0.1Hz
5.000kHz 0.001kHz
50.00kHz 0.01kHz
500.0kHz 0.1kHz
±
(1.2% r 3 digits)
5.000MHz 0.001MHz
±
(1.5% r 4 digits)
Frequency
Sensitivity: >0.5V rms min.
[
1MHz; >3V rms > 1MHz
NOTE:
Accuracy is stated at 65
o
F to 83
o
F (18
o
C to 28
o
C) and less than 75% RH.
Diode Test
Test current of 0.3mA typical, open circuit voltage 1.5V DC typical
Continuity Check
Audible signal will sound if the resistance is less than 150
Ω
(approx.), test current <0.3mA
Temperature Sensor
Requires type K thermocouple
Input Impedance
7.8M
Ω
AC Response
Average responding
ACV Bandwidth
50Hz to 400Hz
Display
4,000 count liquid crystal
Overrange indication
“
OL
” is displayed
Auto Power Off
30 minutes (approximately)
Polarity
Automatic (no indication for positive); Minus (-) sign for negative
Measurement Rate
2 times per second, nominal
Low Battery Indication “
”
is displayed if battery voltage drops below operating
voltage
Battery
Two AAA batteries
Fuses
mA, µA ranges; 0.5A/660V ceramic fast blow
A range; 10A/250V ceramic fast blow
Operating Temperature
41
°
F to 104
°
F (5
°
C to 40
°
C)
Storage Temperature
-4
°
F to 140
°
F (-20
°
C to 60
°
C)
Operating Humidity
Max 80% up to 87
°
F (31
°
C) decreasing linearly to 50% at 104
°
F
(40
°
C)
Storage Humidity
<80%
Operating Altitude
7000ft. (2000meters) maximum.
Weight
0.5lb (225g) (includes holster).
Size
5.31” x 2.67” x 1.7” (135 x 68 x 43mm)
Safety
This meter is intended for indoor use and protected, against the
users, by double insulation per EN61010-1 and IEC61010-1 2nd
Edition (2001) to CAT II 1000V & CAT III 600V; Pollution Degree 2.
The meter also meets UL 61010-1, Second Edition (2004),
CAN/CSA C22.2 No. 61010-1, Second Edition (2004), and UL
61010B-2-031, First Edition (2003)
Approvals
UL
CE
UL Listed
The UL mark does not indicate that this product has been
evaluated for the accuracy of its readings.