NTRODUCTION AND TECHNICAL 

D

ATA

 

The Polyester junction box is designed for the parallel connection of load cells. There are 4 different types: 

 

Type 

No. of load cells 

Housing size 

Inputs 

Output 

KPF-4 

up to 4 

100 x 320 x 81 mm 

4x M16 or 4x M12 

1x M16 

KPF-6 

up to 6 

100 x 320 x 81 mm 

6x M16 or 6x M12 

1x M16 

KPF-8 

up to 8 

100 x 320 x 81 mm 

8x M16 or 8x M12 

1x M16 

KPF-10 

up to 10 

100 x 360 x 90 mm 

10x M16 

1x M16 

 

The junction box type KPF is special designed for weigh bridges and silos including surge arrestors for 
lightning protection. It can be connected to the instrumentation with a shielded 6-wire signal cable. 
The corner correction is done with resistors. Each junction box already contains 2.1 Ohm resistors for each 
load cell from the factory. Therefore it is possible to decrease a load cell signal (use a higher resistance 
value) or to increase a load cell signal (use a smaller resistance value). 

 

Housing material: 

Polyester 

Protection class: 

IP66 

Cable connection: 

With clamping terminals 

Corner correction: 

By exchangeable resistors 

Other: Over-voltage 

protection by surge arrestors 

 

M

ECHANICAL 

I

NSTALLATION

 

Look for a mounting location which is more or less dry and protected from environmental stress. The junction 
box itself has to be mounted with the cable glands showing vertically to the bottom. A flooding or any tensile 
load on the cable glands must be prevented. 
All bottom parts of the cable glands are sealed to the housing with some gaskets. You have to check the 
existence and the condition of these gaskets, the gaskets have to be replaced if required. 
You have to check all bottom parts of the cable glands for a firm seat in the housing. 
(Recommendation: use a 12-point ring spanner with a cut-out and metric size 19 mm, if the cable are already 
installed). 
 

E

LECTRICAL 

C

ONNECTIONS

 

1

2

2

1

1

2

1

2

2

1

 

Figure 1: Dimensions in [mm] 

 

The earth screw (see (2) in fig.1) has to be connected to protective earth or you must establish an electrical 
connection to the weigh modules for potential equalisation. For doing this only the upper nut should be 
loosened and tightened again. The lower nut (= counter nut) and the earth screw with washer must not be 
rotated or loosened – the gasket is bonded! If loosened by accident the earth screw must be re-inserted and 
locked with some Loctite. 
 
The connection sequence of the load cells should correspond to the corners of the scale, i.e 
Corner 1 = Load cell 1, Corner 2 = Load cell 2, etc. 

Junction Box Type KPF – Technical Manual, Rev. 1.11   July 2008

 

Page 4 of 8

Содержание KPF-4

Страница 1: ...Junction Box Type KPF 4 KPF 6 KPF 8 KPF 10 T TE EC CH HN NI IC CA AL L M MA AN NU UA AL L Flintec GmbH Bemannsbruch 9 74909 Meckesheim GERMANY www flintec com...

Страница 2: ...the use of any options or any consumable products other than those designated as Original FLINTEC Products NOTICE The contents of this manual are subject to change without notice Copyright 2007 by Fli...

Страница 3: ...4 MECHANICAL INSTALLATION 4 ELECTRICAL CONNECTIONS 4 LOAD CELL CABLE CONNECTION 5 OUTPUT CABLE CONNECTION 5 CORNER CORRECTION AT SCALES WITH FLINTEC LOAD CELLS 6 CLOSING THE JUNCTION BOX 6 SUMMARY OF...

Страница 4: ...h is more or less dry and protected from environmental stress The junction box itself has to be mounted with the cable glands showing vertically to the bottom A flooding or any tensile load on the cab...

Страница 5: ...ube significantly Unused cable inlets must be closed by the delivered blind plugs If the cable gland nuts are correctly tightened the sealing part of the cable gland must constrict the blind plug sign...

Страница 6: ...kg test load 0 1 kg 500 kg x 1100 0 22 Example 2 400 load cell 10 kg corner error with 5000 kg test load 10 kg 5000 kg x 400 0 8 4 Install the correction resistor into the junction box for the corresp...

Страница 7: ...ing hole must be closed by a blind plug and a gasket 6 The earth screw must be electrically connected with protective earth potential equalisation For doing this only the upper nut should be loosened...

Страница 8: ...ke service kit no 1 but without ring spanner SSG19 Service kit no 3 Article no 5200 U3 for load cells with maximum capacity in the range from 7 5 t to 22 5 t with ring spanner Parts list Description A...

Отзывы: