9 Start-up of the transmitter
68
9.3
Initialization
During an initialization (INIT) of the transmitter, the settings in the program branches
Parameter
and
Output Options
and some of the settings in the program branch
Special Funct.
are reset to the default settings of the manufacturer.
For INIT-resistant settings, see annex A.
Proceed as follows to execute an initialization:
• When switching on the transmitter: keep keys BRK and C pressed.
• During the operation of the transmitter: press keys BRK, C and ENTER at the same time. A reset is executed. Release
only key ENTER. Keep keys BRK and C pressed.
After the initialization, the remaining settings of the transmitter can be reset to the default settings and/or the stored mea-
sured values can be deleted.
9.4
Display
9.4.1
Main menu
The selected program branch is displayed in angle brackets and capital letters. The complete designation of the selected
program branch is displayed in the lower line.
Select a program branch with key
and
. Press ENTER.
9.4.2
Program branches
•
program branch
Parameter
input of the pipe and fluid parameters
•
program branch
Measuring
execution of measurement steps
•
program branch
Output Options
setting of the physical quantity, the unit of measurement and the parameters for the transmission of measured value
•
program branch
Special Funct.
contains all functions that are not directly related to the measurement
After the initialization has been executed, the message
INITIALISATION
DONE
is dis-
played.
Select
yes
to reset the remaining settings to default settings or
no
to keep them at the cur-
rent settings.
Press ENTER.
If
yes
is selected, the message
FACTORY DEFAULT
DONE
will be displayed.
Select
yes
to delete the stored measured values or
no
to keep them. Press ENTER.
This display will only be indicated if measured values are stored in the transmitter.
The main menu contains the following program branches:
•
par
(parameters)
•
mea
(measuring)
•
opt
(output options)
•
sf
(special functions)
Note!
By pressing key BRK, the measurement is stopped and the main menu is selected.
Note!
In this user manual, all program entries are displayed in typewriter font (
Parameter
). The menu
items are separated from the main menu by a backslash "\".
INITIALISATION
----DONE----
FACTORY DEFAULT?
no
>YES<
Delete Meas.Val.
no
>YES<
>PAR< mea opt sf
Parameter
4
6
Содержание FLUXUS WW
Страница 1: ...User manual UMFLUXUS_WWV1 1EN Ultrasonic flowmeter for water and wastewater FLUXUS WW...
Страница 4: ...FLUXUS WW UMFLUXUS_WWV1 1EN 2018 02 28...
Страница 8: ...FLUXUS WW Table of contents 8 UMFLUXUS_WWV1 1EN 2018 02 28...
Страница 144: ...FLUXUS WW D Conformity declarations 144 UMFLUXUS_WWV1 1EN 2018 02 28 D Conformity declarations...
Страница 145: ......
Страница 146: ......