16 Outputs
114
16.1.3
Function test
Connect a multimeter to the installed output.
Test of the analog outputs
Test of the binary outputs
16.2
Error value delay
The error value delay is the time interval after which the entered error value is transmitted to the output in case no valid
measured values are available. The error value delay can be entered in the program branch
Output Options
if this
menu item has previously been activated in the program branch
Special Funct.
. If the error value delay is not entered,
the damping factor will be used.
The current output is tested. Enter a test value. It has to be within the output range. Press
ENTER.
If the multimeter displays the entered value, the output functions correctly.
Select
yes
to repeat the test,
no
to return to
SYSTEM settings
. Press ENTER.
Select
Opto-Relay OFF
in the scroll list
Output Test
to test the de-energized state of
the output. Press ENTER. Measure the resistance at the output. The value has to be high
ohmic.
Select
yes
. Press ENTER.
Select
Opto-Relay ON
in the scroll list
Output Test
to test the energized state of the
output. Press ENTER. Measure the resistance at the output. The value has to be low ohm-
ic.
Select
yes
to repeat the test,
no
to return to
SYSTEM settings
. Press ENTER.
Select
Special Funct.\SYSTEM settings\Dialogs/Menus\Error-val. delay
.
Select
damping
if the damping factor is to be used as the error value delay. Select
edit
to
activate the menu item
Error-val. delay
in the program branch
Output Options
.
From now on, the error value delay can be entered in the program branch
Output Op-
tions
.
I1:Output Test
4
mA
I1= 4.0 mA
Again? no
>YES<
B1:Output Test
↕
Opto-Relay OFF
B1=OFF
Again? no
>YES<
B1:Output Test
↕
Opto-Relay ON
B1=ON
Again? no
>YES<
Error-val. delay
>DAMPING<
edit
Error-val. delay
10
s
Содержание FLUXUS WW
Страница 1: ...User manual UMFLUXUS_WWV1 1EN Ultrasonic flowmeter for water and wastewater FLUXUS WW...
Страница 4: ...FLUXUS WW UMFLUXUS_WWV1 1EN 2018 02 28...
Страница 8: ...FLUXUS WW Table of contents 8 UMFLUXUS_WWV1 1EN 2018 02 28...
Страница 144: ...FLUXUS WW D Conformity declarations 144 UMFLUXUS_WWV1 1EN 2018 02 28 D Conformity declarations...
Страница 145: ......
Страница 146: ......