A Menu structure
125
Annex
A
Menu structure
INIT-resis-
tant
Program branch
Parameter
main menu: selection of the program branch
Parameter
selection of a measuring channel (A, B) or a calculation channel (Y, Z)
This display will not be indicated if the transmitter has only one measuring
channel.
When a measuring channel is selected (A, B)
input of the outer pipe diameter
input of the pipe circumference
This display will only be indicated if
Special Funct.\SYSTEM set-
tings\Dialogs/Menus\Pipe Circumfer.
is activated and
Outer
Diameter
=
0
has been entered.
input of the pipe wall thickness
range: depends on the connected transducers
default: 3 mm
selection of the pipe material
input of the sound speed of the pipe material
range: 600…6 553.5 m/s
This display will only be indicated if
Other Material
is selected.
selection whether the pipe is lined
selection of the lining material
This display will only be indicated if
Lining
=
yes
is selected.
input of the sound speed of the lining material
range: 600…6 553.5 m/s
This display will only be indicated if
Other Material
is selected.
input of the liner thickness
range: 0...50 mm
default: 3 mm
input of the roughness of the inner pipe wall
range: 0…5 mm
default: 0.1 mm (for steel as pipe material)
input of the fluid temperature
default: 20 °C
>PAR< mea opt sf
Parameter
Parameter
↕
for Channel
A:
Outer Diameter
100.0
mm
Pipe Circumfer.
314.2
mm
Wall Thickness
3.0
mm
Pipe Material
↕
Carbon Steel
c-Material
3230.0
m/s
Lining
no
>YES<
Lining
↕
Bitumen
c-Material
3200.0
m/s
Liner Thickness
3.0
mm
Roughness
0.4
mm
Medium Temperat.
20.0
C
Содержание FLUXUS WW
Страница 1: ...User manual UMFLUXUS_WWV1 1EN Ultrasonic flowmeter for water and wastewater FLUXUS WW...
Страница 4: ...FLUXUS WW UMFLUXUS_WWV1 1EN 2018 02 28...
Страница 8: ...FLUXUS WW Table of contents 8 UMFLUXUS_WWV1 1EN 2018 02 28...
Страница 144: ...FLUXUS WW D Conformity declarations 144 UMFLUXUS_WWV1 1EN 2018 02 28 D Conformity declarations...
Страница 145: ......
Страница 146: ......