9 Measurement
FLUXUS F501
9.1 Parameter input
71
UMFLUXUS_F501V1-3EN, 2020-06-30
Fluid density
The density is used to calculate the mass flow.
If the mass flow is not measured, an input is unnecessary. The default value can be used.
• Enter the operating density of the fluid.
• Press ENTER.
This display will only be indicated if
Other Medium
is selected.
Fluid temperature
At the beginning of the measurement, the fluid temperature is used for the interpolation of the sound speed and therefore
for the calculation of the recommended transducer distance.
During the measurement, the fluid temperature is used for the interpolation of the density and viscosity of the fluid.
The value entered here is used for the calculation if the fluid temperature is not measured.
• Enter the fluid temperature. The value has to be within the operating temperature of the transducers.
• Press ENTER.
9.1.3
Other parameters
Transducer parameter
If transducers are identified at a measuring channel, the input of parameters is finished.
• Press ENTER. The main menu will be displayed.
If no or special transducers are connected, the transducer parameters have to be entered.
• Select
Standard
to use the standard transducer parameters stored in the transmitter.
• Select
Special Version
to enter the transducer parameters. The transducer parameters have to be provided by
FLEXIM.
• Press ENTER.
If
Special Version
is selected, enter the 6 transducer parameters specified by FLEXIM. Press ENTER after each input.
Extension cable
In case the transducer cable is extended (e.g., between junction box and transmitter), enter the length of the extension cable.
• Select the menu item
Additional cable
in the program branch
Parameter
.
• Enter the length of the extension cable.
• Press ENTER.
Parameter\Medium\Other Medium\Density
Parameter\Medium Temperat.
Parameter\Transducer Type
Notice!
If a standard transducer is selected, no transducer-specific calibration values are considered. A higher uncertainty
has to be expected.
Parameter\Transducer Type\Special Version
Parameter\Additional cable
Содержание FLUXUS F501
Страница 1: ...FLUXUS F501 UMFLUXUS_F501V1 3EN Operating instruction...
Страница 6: ...Table of contents FLUXUS F501 2020 06 30 UMFLUXUSF501V1 3EN 6...
Страница 143: ...Annex FLUXUS F501 D Conformity declarations 143 UMFLUXUS_F501V1 3EN 2020 06 30 D Conformity declarations...
Страница 144: ...Annex D Conformity declarations FLUXUS F501 2020 06 30 UMFLUXUS_F501V1 3EN 144...
Страница 145: ......
Страница 146: ......