Supplement to User Manual - ENGLISH
62
Request of the State of Multiple Measured Values
For the start addresses and the offsets see Tab. 7.3 and Tab. 7.4.
The state addresses are references to the state registers from Tab. 6.3.
heat flow rate, totalizer for positive flow direction
34
35
heat flow rate, totalizer for negative flow direction
36
37
concentration
38
39
medium temperature (supply line)
40
41
medium temperature (return line)
42
43
medium pressure (supply line)
44
45
medium pressure (return line)
46
47
compressibility coefficient of the gas (gas measurement)
48
49
density
50
51
volumetric flow rate (net),
difference between the totalizers of the positive and negative flow direction
52
53
standard volumetric flow rate (net),
difference between the totalizers of the positive and negative flow direction
54
55
mass flow rate (net),
difference between the totalizers of the positive and negative flow direction
56
57
heat flow rate (net)
difference between the totalizers of the positive and negative flow direction
58
59
Example:
measuring channel: B
physical quantities: medium temperature (supply line),
medium temperature (return line),
medium pressure (supply line),
medium pressure (return line)
Modbus register start address: 16000 + 40 = 16040
Modbus register range: 16040...16047
register number: 8
Tab. 7.3: Start addresses of the channels
channel
start address
A
15200
B
16200
C
17200
D
18200
Z
19200
Y
20200
Tab. 7.2: Offset of the measured value
physical quantity (floating point number 32 bit)
offset
lo word
hi word
Содержание FLUXUS ADM 5 07 Series
Страница 4: ...Modbus Server Supplement to User Manual 4 SU_ModbusV2 8 2015 03 02...
Страница 6: ...Modbus Server Erg nzung zur Bedienungsanleitung DEUTSCH 6 SU_ModbusV2 8 2015 03 02...
Страница 46: ...Modbus Server Erg nzung zur Bedienungsanleitung DEUTSCH 46 SU_ModbusV2 8 2015 03 02...
Страница 48: ...Modbus Server Supplement to User Manual ENGLISH 48 SU_ModbusV2 8 2015 03 02...