![Flexicon 61-110-014 Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/flexicon/61-110-014/61-110-014_operators-manual_2304652013.webp)
Operators and Reference Manual
MC12
MC12 OM RM EN 74-111-132 v1.03.doc
Ver. 1.03
Page 13 of 47
6.7 Description of functions
The individual functions will be described in the following:
6.7.1. Function 1 -Volume
Value: Choice of ml and grams as unit.
Function 1 informs the system of the volume to be filled.
Value
Option
Related Function(s)
Min
Max
0.01
9999.9
ml. or gram
72
The entered value must be between 0.01 and 9999.9. If a value outside these limits is
entered, the MC12 will correct the value to the nearest value within the limits and ask for
acceptance via <ENT>.
When a volume is entered, the system will calculate the new parameter value which will then
be used from the start of the next fill. At the same time, the status line will be updated.
For an individual fill only the parameter value for the drive in question will be changed.
For parallel or serial filling, the volume should be entered in drive 0 and the parameter value
will be calculated for all connected drives.
6.7.2. Function 2 - Tube diameter
Value: Inside diameter of the tube in mm.
Drive type
Tube inside diameter in mm
PD12P
0.5 0.8 1.2 1.6 3.2 4.8 6.0 8.0
Enter the tube diameter used. The MC12 will only accept original Flexicon tube diameters.
The diameters are specified in the manual for the drive and can be read on the tube gauge
supplied with the drive.
As most of Flexicons drives are of the peristaltic type and work with tubes, the system must
know the current tube diameter.
When programming a drive that does not use tubes as pumping element, this function will
not be activated.
The tubes should be indicated by their inside diameter and from these data the MC12 will
calculate how many revolutions the selected pump head should make in order to dispense
the required volume.
If an unrecognised tube diameter is entered for the selected drive, the MC12 will not accept
this diameter but will instead ask for a tube diameter from the original Flexicon range.