15.0 Tube 

Replacement

MAINTENANCE MODE

Pump Cover Removed!

Press STOP to stop

- Cover was removed!
- Motor Speed = 6 RPM

MAINTENANCE MODE

Press START to run

Pump Head

Spacer, Rear

Rotor

Spacer, Front

Tubing Assembly

Pump Head Cover

Thumb
Screws

Tube Nut

Remove 

Pump Head Cover 

by unscrewing four

 Thumb Screws. 

Pull out

 Pump 

Head Cover. 

Pump will detect 

Pump Head Cover

 is removed and enter MAINTENANCE MODE.

Rotor will rotate at a maximum of 6 RPM for your safety.

Pull out suction side of 

Tubing Assembly

.

Press START

 

button. While rotor is rotating, pull out old 

Tube Assembly

.

TIP! Let pump do the work for you. Just guide tubing out between two rollers 
located on 

Rotor

.

Press STOP button at any time to stop the pump.

Pull out suction line adapter from Pump Head. Pull out 

Tubing Assembly

 as the 

Rotor

 rotates around.

Stop pump by pressing STOP button.

Thoroughly clean 

Pump Head

 and 

Rotor

Rotor

 can be removed by pulling straight 

out. After cleaning process, push 

Rotor

 back on shaft. See drawing above for 

proper assembly. IMPORTANT! 

Rotor

 direction; the word “FRONT” on 

Rotor

 must 

face front of pump.

Locate your new tubing and Tube Installation Tool. Please see next page on how to 
install new 

Tube Assembly

 into 

Pump Head

.

!

CAUTION

Prior to service, pump clean water through the pump and suction / discharge line to remove 
chemical.

Always wear protective clothing, face shield, safety glasses and gloves when working on 
or near your metering pump.  Additional precautions should be taken depending on 
solution being pumped. Refer to MSDS precautions from your solution supplier.

!

CAUTION

Tube Installation Tool

90002-278

Use provided Tube Installation Tool to leverage tubing into pump head, NOT YOUR 
FINGERS.

!

CAUTION

Use extreme caution when replacing pump tube. Be careful of your fingers and DO NOT 
place fingers near rollers.

!

CAUTION

Flex-Pro

Содержание Norprene A3

Страница 1: ...e co id or info spe co id URL www spe co id SERIES A3V and A4V Operating Manual ProSeries by Blue White Ind TM Patents 7 001 153 7 284 964 4 496 295 and other patents pending FLEX PRO Peristaltic Mete...

Страница 2: ...onal Dosing 24 10 0 Contact Closure Input 25 11 0 Set FVS flow verification system 26 12 0 Output Setup Signal Relay Contact Closure 27 12 1 Signal Output 28 13 0 Pump Tube Timer 29 14 0 TFD tube fail...

Страница 3: ...C 130 54 130 54 F C 130 54 130 54 130 54 130 54 F C 185 85 185 85 185 85 185 85 185 85 Norprene A3 Tube Pumps Meets FDA criteria for food Excellent chemical resistance Extra long life at low pressures...

Страница 4: ...15 1 4 D 20 6W x 27 3H x 38 9D cm A4 models 12 1 8 W x 14 1 4 H x 18 5 8 D 30 8W x 36 1H x 47 3D cm Approximate shipping wt A3 models 33 lb 15 0 Kg A4 models 56 lb 25 4 Kg Wetted components Non Wette...

Страница 5: ...15V 1A contact closures assignable to monitor up to 17 different pump functions including TFD FVS remote local control setting motor on fault current operating mode and others Two CNC precision machin...

Страница 6: ...nly n Raise metering pump 4 1 2 inches 11 43 cm off ground or a surface n Made out of tough Stainless Steel n Provides a stable mounting surface Stainless Steel extended brackets allow pump to be secu...

Страница 7: ...buildup when injecting chemicals that calcify such as sodium hypochlorite Duckbill may add additional back pressure to pump up to 7 psi 3 8 OD x 1 4 ID Tubing and Tri Clamp connection Models Proper ey...

Страница 8: ...and power cables together Power Interruption This pump has an auto restart feature which will restore pump to operating state it was in when power was lost Note When in doubt regarding your electrical...

Страница 9: ...CE ACTIVE LOOP OPEN COLLECTOR 0 1000 Hz 50 DUTY CYCLE 6 AMP MAX AT 250 VAC 5 AMP MAX AT 30 VOLT DC 1 AMP MAX AT 125 VOLT AC 0 8 AMP MAX AT 30 VOLT DC 1 AMP MAX AT 125 VOLT AC 0 8 AMP MAX AT 30 VOLT DC...

Страница 10: ...a runtime function and is also used when configuring your pump Runtime Function While pump is operating in any Run Mode you can view useful information regarding your current Run Mode by pressing RIG...

Страница 11: ...until STOP or START button is pressed STOP button will stop pump and display will read OFF START button will stop prime feature and jump back into previous operating state To REVERSE ROTATION of rotor...

Страница 12: ...ist of all available options Scroll bar example While making a selection where only one choice is allowed you will see a radio button Radio button example In a screen where you are making changes you...

Страница 13: ...lay Language Esc Reset to Defaults Set Password Display Rate Reset to Factory Defaults Reset to Factory Defaults Reset to Factory Defaults Are you sure Yes No Liter Hr Gal Hr US Gal day US Done Displa...

Страница 14: ...Display Rate Press MENU button to enter menu structure for setting up pump Select Configuration and Press ENTER button Select Display Rate and Press ENTER button Select your desired Display Rate unit...

Страница 15: ...fault setting is DISABLE Set FVS Flow Verification System Set Flow Verification time delay Use this feature if you are using a Blue White flow verification sensor to monitor flow output Default settin...

Страница 16: ...nu structure To do this you must select Esc at bottom of every screen menu until you see Run Mode screen displayed Tip This feature can be combined with Contact Input feature to allow for remote Start...

Страница 17: ...t Esc at bottom of every screen menu until you see Run Mode screen displayed 4 20 mA Input Motor Min Signal Input Motor Speed at Minimum mA Signal 00 0 0 4 20 mA Input Min mA Signal Input Min mA signa...

Страница 18: ...ture To do this you must select Esc at bottom of every screen menu until you see Run Mode screen displayed 0 10 VDC Input Motor Min Signal Input Motor Speed at Minimum Volt DC Signal 00 0 0 0 10 VDC I...

Страница 19: ...ency Input setup you must navigate back out of menu structure To do this you must select Esc at bottom of every screen menu until you see Run Mode screen displayed Frequency Input Motor Min Signal Inp...

Страница 20: ...hold Select Run Mode button until Run Mode screen is displayed If you used Menu button to navigate to Pulse Batch setup you must navigate back out of menu structure To do this you must select Esc at...

Страница 21: ...setup then just press and hold Select Run Mode button until Run Mode screen is displayed If you used Menu button to navigate to Cycle Adjustment setup you must navigate back out of menu structure To...

Страница 22: ...arrow to scroll through 0 9 on selected digit Press RIGHT arrow to scroll over to next digit to right If you pass your desired digit you can easily scroll back by continuously pressing RIGHT button P...

Страница 23: ...desired digit you can easily scroll back by continuously pressing RIGHT button Press ENTER to save changes Continue this process until all four screens have been configured If you used short cut to e...

Страница 24: ...easily scroll back by continuously pressing RIGHT button Press ENTER to save the changes Continue this process until all four screens have been configured If you used short cut to enter Proportional D...

Страница 25: ...croll down to Done selection Then press ENTER To navigate back out of menus select Esc and press the ENTER button at bottom of every screen menu until you see Run Mode screen displayed IMPORTANT If Co...

Страница 26: ...d during stopper operation such as with chlorine an alarm delay time value from 1 255 seconds must be programmed An alarm delay value of 000 seconds disables FVS system If FVS alarm occurs pump will s...

Страница 27: ...ontact 1 Contact 2 Contact 3 Esc Relay Output 4 20 mA Output Signal min Motor Input mA Signal Min Motor speed 4 20 mA Output Minimum Motor Input Min Motor for mA Signal Output 4 20 mA Output Signal Ma...

Страница 28: ...mA 100 Pump Output 20 mA Signal Output Frequency Output Disable Done 4 20 mA Output Default 4 20 mA output settings shown Signal Output Disable Done 4 20 mA Output Default Frequency Hz output settings...

Страница 29: ...14 0 TFD Tube Failure Detection Flex Pro is equipped with a Tube Failure Detection System which is designed to stop pump and provide an output alarm see Output menu in the event pump tube should rupt...

Страница 30: ...Stop pump by pressing STOP button Thoroughly clean Pump Head and Rotor Rotor can be removed by pulling straight out After cleaning process push Rotor back on shaft See drawing above for proper assembl...

Страница 31: ...to slide discharge fitting into pump head tongue and groove Once discharge fitting is secured in pump head stop pump by pressing STOP button Replace pump head cover Pump will ask you if you d like to...

Страница 32: ...Page 32 This Page Blank Flex Pro...

Страница 33: ...Page 33 This Page Blank Flex Pro...

Страница 34: ...Page 34 This Page Blank Flex Pro...

Страница 35: ...to allow back flow If your desire is to simply extend tube life Typically tubing fails on outlet side pressure side of tube assembly in pump head Failure to install checkvalves in their proper flow d...

Страница 36: ...mical from pump Determine which parts are being attacked and replace them with parts that have been manufactured using more suitable materials Manufacturer does not assume responsibility for damage to...

Страница 37: ...ller assemblies rotors in this group are same size and interchangeable Tubing in this group are interchangeable Norprene Group All roller assemblies rotors in this group are same size and interchangea...

Страница 38: ...Tubing ID Example Tube Assembly reorder number with NH printing A3 SNH T A 3 S N H T Pump Series Tubing Material N Norprene T Norprene Chemical G Tygothane Tubing ID D 075 E 125 F 155 G 187 H 250 J 3...

Страница 39: ...tested with water only for pressure and performance Installers and operators of these devices must be well informed and aware of the precautions to be taken when injecting various chemicals especiall...

Отзывы: