11.0 Set FVS (Flow Verification System)
Used to monitor pump output. If pump does not dispense fluid when pump
head rotor is turning, pump will go into an alarm mode and stop pump. Blue-
White offers a flow verification sensor that can easily attach to fitting on pump.
Default settings:
000 (off)
Navigate to
Set FVS
menu by pressing MENU button, then selecting Input
Setup, and then
Set FVS
.
Press UP or DOWN arrow to scroll through 0 - 9 on selected digit.
Press RIGHT arrow to scroll over to next digit to right. If you pass your desired
digit, you can easily scroll back by continuously pressing RIGHT button.
Press ENTER to save changes and exit FVS screen.
To navigate back out of menus, select <-Esc and press ENTER button at
bottom of every screen menu until you see Run Mode screen displayed.
Flow Verification System
(sensor sold separately)
A3 is equipped with a
Flow Verification System
which is designed to stop pump
and provide a contact closure output or 10 amp relay output (setup in software)
in event sensor does not detect chemical during pump operation. This could
indicate a clogged injection fitting, empty chemical solution tank, worn pump
tube, loose tubing connection, etc.
To allow pump to clear any gasses that may have accumulated during stopper
operation (such as with chlorine), an alarm delay time value from 1-255
seconds must be programmed (An alarm delay value of 000 seconds disables
FVS system).
If FVS alarm occurs, pump will stop, send an external signal (if setup), and
screen will flash FVS with an alarm icon.
To clear FVS alarm, you must press START button or re-cycle power (unplug
power to pump, then plug back in).
Install FVS Flow Sensor -
Flow Verification Sensor (FVS) should be installed
on the inlet (suction) side of the pump tube. Sensor includes a PVC tubing
insert, located inside sensors female thread connection, that is designed to
seal sensor onto pump tube inlet adapter. Thread sensor onto pump tube until
tubing insert is snug against pump tube inlet fitting - do not over-tighten.
- Set FVS
- Input Setup
+ Set FVS
<- Esc
Hidden
+ Contact Closure Input
+ Motor - Select RPM
+ Tubing - Max Flowrate
+ Operating Input Modes
Input FVS Time Delay:
OFF = 000
Range = 001-255 seconds
00
0
Seconds
Confirm the FVS flow range -
Flow
Verification Sensor (FVS) will only function
within its operating range. Sensor model
FV-100-6V has an operating range of 30-
300 ml/min (1-10 oz/min). If the pump’s
output is less than 30 ml/min (0.5 ml/sec),
sensor will not detect chemical and a
signal will not be sent to pump.
OPERATING
FLOW RANGE
(ml/min)
100-1000
FV-200-6V
200-2000
FV-300-6V
300-3000
FV-400-6V
500-5000
FV-500-6V
700-7000
FV-600-6V
SENSOR
MODEL
NUMBER
FVS
ALARM
When a FVS alarm occurs
STOP