Chapter 2 | Getting Started 

2.3 Keypad Functions 

 

  

Key 

Function 

 

In Continuous measure mode, press to hold 

a reading on the display and then press 

again to release a reading on the display. 

In Auto-Read measure mode, press to start 

a new reading.  

Press to escape the Setup Menu and 

Calibration mode without saving changes. 

 

In Continuous measure mode, press to save 

the reading to the data log and export to an 

external device if one is connected. 

 

Press to enter the Setup Menu.  

Press to exit the Setup Menu. 

 

Press to enter the Calibration mode.  

Press while in the Calibration mode to save 

and end the calibration when performing a 

one to four point pH calibration. 

 

Press to scroll through menu items in the 

Setup Menu; the list is cyclical.  

In the Measure mode, press and hold (long 

press) for three seconds as a shortcut key to 

change the measure type from pH to RmV. 

 

Press to change a setting in the Setup 

Menu. Press to scroll through a list or 

change a numeric value. 

In the Calibration mode, press to manually 

change the calibration value. 

In the Measure mode, press and hold (long 

press) for three seconds as a shortcut key to 

view current date and time setting.  

 

Press to save a setting in the Setup Menu. 

Press to accept a calibration point in the 

Calibration mode.  

 

Press to power on the meter. When the 

meter is on, press to turn the backlight off or 

on. When the meter is on, press and hold 

for three seconds to power off the meter. 

2.3.1 Setup Menu Scrolling 

To scroll through the Setup Menu list: 

 

Press the “Menu” key to enter the Setup Menu. 

 

The first Setup Menu item will always be show. 

 

Press the 

 key to scroll to the second item or 

press the 

 key to scroll to the last item.  

 

The item list is cyclical and will scroll from the last 
item to the first item as the 

 key is pressed. 

2.3.2 Setup Menu Value Changes 

For all numeric value changes, press the 

 or 

 key to 

edit a value: 

 

Press the 

 key once to increase the value by one 

least significant digit/unit. 

 

Press the 

 key once to decrease the value by one 

least significant digit/unit. 

 

Press and hold the 

 key to quickly increase the 

numeric value. 

 

Press and hold the 

 key to quickly decrease the 

numeric value. 

2.3.3 Mode Shortcut Key 

In the main measure mode, press and hold (long press) the 

 or 

 key for approximately three seconds to change the 

main measure mode to pH or relative mV (RmV). 

2.3.4 Date and Time View 
Shortcut Key 

In the main measure mode, press and hold (long press) the 

 or 

 key for approximately three seconds to view the 

current date and time settings.  

Содержание accumet AB315

Страница 1: ...Operations Manual Electrochemistry Meters 68X002124 Revision 01 September 2021 accumet AB315 pH mV Meter User Manual...

Страница 2: ......

Страница 3: ...3 1 Setup Menu Scrolling 8 2 3 2 Setup Menu Value Changes 8 2 3 3 Mode Shortcut Key 8 2 3 4 Date and Time View Shortcut Key 8 2 4 Display Overview 9 2 5 Meter Connections 10 2 6 Electrode Information...

Страница 4: ...19 3 2 16 Sleep Mode 19 3 2 17 Clear Data 19 3 2 18 Factory Reset 20 Chapter 4 21 Calibration 21 4 1 pH Calibration 21 4 1 1 pH Calibration Overview 21 4 1 2 pH Calibration Procedure 21 4 1 3 pH Calib...

Страница 5: ...28 7 2 Meter Dimensions 29 7 3 Warranty 29 7 3 1 Return of Items 29 Chapter 8 30 Regulatory Compliance 30 8 1 European Union 30 8 2 Product Safety 30 8 3 Electromagnetic Compatibility 30 8 3 1 FCC Sta...

Страница 6: ...ol in the context of a Note indicates to follow the manufacturer specified instructions notes and requirements set out in the instruction manuals CAUTION Before using or maintaining this product pleas...

Страница 7: ...US Japan Euro UK Singapore Australia New Zealand and China wall plug plates is included with the meter This universal power adapter is specifically for use with this meter Use of other power adapters...

Страница 8: ...etting Press to save a setting in the Setup Menu Press to accept a calibration point in the Calibration mode Press to power on the meter When the meter is on press to turn the backlight off or on When...

Страница 9: ...Icon Indicates meter is in the Setup Menu and viewing the data log or calibration log Log Saved Icon Indicates meter is actively saving data to the data log or calibration log Continuous Icon Indicate...

Страница 10: ...separately or kitted with different electrodes Refer to the electrode instructions for specific care and maintenance procedures The following are general recommendations for electrode preparation Remo...

Страница 11: ...d data logs and calibration logs to the selected export device 4 Temperature Calibration or Manual Temperature Input When an ATC probe is connected use the Temp Calibration menu to perform a temperatu...

Страница 12: ...key to view the display log 3 2 1a Viewing the Data Log Press the Enter key to view the display log For the data log press the or key to scroll through the data log points Scroll through data log poin...

Страница 13: ...calibration log data Press key Press key Press key Press key Press key If desired press the Log Export key to export the individual calibration log being viewed Press the Menu key to return to the mai...

Страница 14: ...he Setup Menu list until the Temp Calibration item is shown The temperature value will blink indicating the reading from the ATC probe is being measured Once the reading is stable the value will stop...

Страница 15: ...ress the Enter key to save the change 3 2 6 pH Resolution Set the displayed pH resolution as 0 1 or 0 01 pH units Press the or key to scroll through the Setup Menu list until the pH Resolution item is...

Страница 16: ...d In the Timed measure mode the measurement values are continuously updated on the display Measurements are automatically saved to the data log and exported at the selected time intervals from 5 secon...

Страница 17: ...ly increase b Press the key to decrease the alarm value by one hour Press and hold the key to quickly decrease Press the Enter key to save the change 3 2 11 Temperature Units Set the temperature units...

Страница 18: ...and hold the key to quickly increase b Press the key to decrease by one day Press and hold the key to quickly decrease Press the Enter key to save the change The year value will flash Press the or ke...

Страница 19: ...the or key to set the audio mode Press the Enter key to save the change 3 2 16 Sleep Mode Set the sleep mode on or off When the meter sleep mode is on the meter will enter sleep mode when no keys are...

Страница 20: ...ta logs calibration logs and settings and return the meter to its factory default state Press the or key to scroll through the Setup Menu list until the Factory Reset item is shown Press the or key to...

Страница 21: ...s to be manually adjusted use the or key to edit the value of the pH buffer at its measured temperature To eliminate temperature errors associated with the pH electrode use an automatic temperature co...

Страница 22: ...ion The active Calibration Buffer Set is shown Rinse the pH electrode and ATC probe and place into the pH 4 01 buffer Wait for the pH value to stabilize Once the reading is stable press the Enter key...

Страница 23: ...11 12 14 40 1 69 4 03 6 84 9 07 11 99 50 1 71 4 06 6 83 9 01 11 73 60 1 72 4 09 6 84 8 96 11 50 70 1 74 4 12 6 85 8 92 11 30 80 1 77 4 16 6 86 8 89 11 13 DIN Buffer Set Temp C pH 1 09 Buffer pH 3 06...

Страница 24: ...Calibration The thermistor sensor used for automatic temperature compensation and measurement is both accurate and stable so frequent calibration is not required When an ATC probe is connected to the...

Страница 25: ...are automatically saved to the data log and exported at the selected time intervals from 5 seconds to 60 minutes the entire time the meter is in the measurement mode 5 2 pH and ORP Measurements Rinse...

Страница 26: ...Baud Rate 9600 bps Data bits 8 Parity None Stop bits 1 Flow Control None 5 5 1 Remote Commands Remote commands allow the meter to be interfaced with computer software like LIMs and HyperTerminal The r...

Страница 27: ...pH buffers and pH electrode o Replace pH buffers o Clean and condition pH electrode o Replace pH electrode as needed 6 2 Measurement Troubleshooting Reading is unstable slow to stabilize Clean electr...

Страница 28: ...t Up to 5 C up to 9 F Features Display 5 backlit LCD Time and Date Yes Read Types Continuous Auto Read Timed Timed Interval 5 seconds to 60 minutes Hold Function Yes Data Log 500 data sets with time a...

Страница 29: ...ired on a charged basis The warranty on your meter shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the...

Страница 30: ...ironment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communicat...

Страница 31: ...entsorgt werden Mehr Informationen ber die Einhaltung dieser Anweisungen durch Thermo Scientific dieVerwerter und Hinweise die Ihnen n tzlich sein k nnen die Thermo Fisher Scientific Produkte zu iden...

Страница 32: ...13 620 19 accumet ORP electrode 13 620 81 accumet meter attached electrode stand 13 637 671 accumet AB315 and AB330 meter replacement universal power cord 110 240V 13 637 010 accumet AB315 and AB330...

Отзывы: