accumet AB315 pH/mV Meter 15
Press the
or
key to adjust the temperature
value.
a.
Press the
key to increase the value by 0.1°.
Press and hold the
key to quickly increase.
b.
Press the
key to decrease the value by 0.1°.
Press and hold the
key to quickly decrease.
Press the “Enter” key to save the change.
3.2.5 Measure Mode
Set the main measure mode to pH or relative mV (RmV).
Press the
or
key to scroll through the Setup
Menu list until the Measure Mode item is shown.
Press the
or
key to set the measure mode.
Press the “Enter” key to save the change.
3.2.6 pH Resolution
Set the displayed pH resolution as 0.1 or 0.01 pH units.
Press the
or
key to scroll through the Setup
Menu list until the pH Resolution item is shown.
Press the
or
key to set the pH resolution.
Press the “Enter” key to save the change.
3.2.7 Calibration Buffer Set
Set the pH buffer set as USA, NIST, DIN or FSCI for
automatic buffer recognition during pH calibrations.
USA: 2.00, 4.01, 7.00, 10.01, 12.00
NIST: 1.68, 4.01, 6.87, 9.18, 12.46
DIN: 1.09, 3.06, 4.65, 6.79, 9.23, 12.75
FSCI: 1.00, 3.00, 6.00, 8.00, 10.00, 13.00
Press the
or
key to scroll through the Setup
Menu list until the Calibration Buffer Set item is
shown.
Press the
or
key to set the pH buffer set.