Fisherbrand accumet AB315 Скачать руководство пользователя страница 25

 

accumet AB315 pH/mV Meter      25 

Chapter 5 
Measurements 

5.1 Read Types 

The 

Read Type

 selected in the meter Setup Menu will 

determine how measurements are performed and when 
measurements are saved to the data log and exported to 
an external device. Set the Read Type as Continuous, 
Auto-Read or Timed to define. 

5.1.1 Continuous  

In the Continuous measure mode, measurements are 
continuously updated on the display and the stopwatch 
(stabilizing) or checkmark (stable) icon indicates the 
measurement stability status.  

Press the “Log/Export” key to save a measurement to the 
data log and export.  

At any time during the measurement, press the “Hold” key 
to lock the measurement on the display (Hold icon will be 
shown on the display) and then press the Hold key a 
second time to release the hold and view live 
measurements again. 

5.1.2 Auto-Read 

In the Auto-Read measure mode, press the “Read” key to 
start a measurement.  

When the measurement is stable, the checkmark (stable) 
icon is shown and the measurement is locked on the 
display until the “Read” key is pressed again.  

The stable measurement is automatically saved to the data 
log and exported.  

5.1.3 Timed 

In the Timed measure mode, the measurement values are 
continuously updated on the display.  

Measurements are automatically saved to the data log and 
exported at the selected time intervals, from 5 seconds to 
60 minutes, the entire time the meter is in the 
measurement mode. 

 

5.2 pH and ORP 

Measurements 

 

Rinse the electrode with distilled water or 
appropriate solution and blot gently with a lint-free 
tissue to remove excess water. 

 

Place the electrode into the sample, submerging 
the measurement sensor and reference junction. 

 

Allow time for the reading to stabilize.  

 

The meter displays the stopwatch icon while 
the reading is stabilizing.  

 

The meter displays the checkmark icon when 
the reading is stable.  

 

Note the displayed measurements as required. 

 

Remove the electrode from the sample. 

 

Repeat steps 1-5 for all samples. 

 

When all samples have been measured, store the 
electrode per instructions in the electrode manual. 

5.3 pH Electrode Icon 

The pH Electrode Icon indicates the condition of the pH 
electrode based on the last saved calibration data and 
resulting slope value.  

 

Electrode Good: pH calibration slope is 92.0% to 102.0% 

Electrode Warning: pH calibration slope is 82.0% to 91.9% 
or 102.1% to 114.0%, recommended actions include: 

 

Check that all pH buffers are fresh, uncontaminated 
and not expired 

 

Clean and condition the pH electrode 

Electrode Error: pH calibration slope is less than 82.0% or 
higher than 114.0%, recommended actions include: 

 

Replace all pH buffers  

 

Replace pH electrode 

Refer to the 

Troubleshooting

 section for more 

recommended actions. 

Содержание accumet AB315

Страница 1: ...Operations Manual Electrochemistry Meters 68X002124 Revision 01 September 2021 accumet AB315 pH mV Meter User Manual...

Страница 2: ......

Страница 3: ...3 1 Setup Menu Scrolling 8 2 3 2 Setup Menu Value Changes 8 2 3 3 Mode Shortcut Key 8 2 3 4 Date and Time View Shortcut Key 8 2 4 Display Overview 9 2 5 Meter Connections 10 2 6 Electrode Information...

Страница 4: ...19 3 2 16 Sleep Mode 19 3 2 17 Clear Data 19 3 2 18 Factory Reset 20 Chapter 4 21 Calibration 21 4 1 pH Calibration 21 4 1 1 pH Calibration Overview 21 4 1 2 pH Calibration Procedure 21 4 1 3 pH Calib...

Страница 5: ...28 7 2 Meter Dimensions 29 7 3 Warranty 29 7 3 1 Return of Items 29 Chapter 8 30 Regulatory Compliance 30 8 1 European Union 30 8 2 Product Safety 30 8 3 Electromagnetic Compatibility 30 8 3 1 FCC Sta...

Страница 6: ...ol in the context of a Note indicates to follow the manufacturer specified instructions notes and requirements set out in the instruction manuals CAUTION Before using or maintaining this product pleas...

Страница 7: ...US Japan Euro UK Singapore Australia New Zealand and China wall plug plates is included with the meter This universal power adapter is specifically for use with this meter Use of other power adapters...

Страница 8: ...etting Press to save a setting in the Setup Menu Press to accept a calibration point in the Calibration mode Press to power on the meter When the meter is on press to turn the backlight off or on When...

Страница 9: ...Icon Indicates meter is in the Setup Menu and viewing the data log or calibration log Log Saved Icon Indicates meter is actively saving data to the data log or calibration log Continuous Icon Indicate...

Страница 10: ...separately or kitted with different electrodes Refer to the electrode instructions for specific care and maintenance procedures The following are general recommendations for electrode preparation Remo...

Страница 11: ...d data logs and calibration logs to the selected export device 4 Temperature Calibration or Manual Temperature Input When an ATC probe is connected use the Temp Calibration menu to perform a temperatu...

Страница 12: ...key to view the display log 3 2 1a Viewing the Data Log Press the Enter key to view the display log For the data log press the or key to scroll through the data log points Scroll through data log poin...

Страница 13: ...calibration log data Press key Press key Press key Press key Press key If desired press the Log Export key to export the individual calibration log being viewed Press the Menu key to return to the mai...

Страница 14: ...he Setup Menu list until the Temp Calibration item is shown The temperature value will blink indicating the reading from the ATC probe is being measured Once the reading is stable the value will stop...

Страница 15: ...ress the Enter key to save the change 3 2 6 pH Resolution Set the displayed pH resolution as 0 1 or 0 01 pH units Press the or key to scroll through the Setup Menu list until the pH Resolution item is...

Страница 16: ...d In the Timed measure mode the measurement values are continuously updated on the display Measurements are automatically saved to the data log and exported at the selected time intervals from 5 secon...

Страница 17: ...ly increase b Press the key to decrease the alarm value by one hour Press and hold the key to quickly decrease Press the Enter key to save the change 3 2 11 Temperature Units Set the temperature units...

Страница 18: ...and hold the key to quickly increase b Press the key to decrease by one day Press and hold the key to quickly decrease Press the Enter key to save the change The year value will flash Press the or ke...

Страница 19: ...the or key to set the audio mode Press the Enter key to save the change 3 2 16 Sleep Mode Set the sleep mode on or off When the meter sleep mode is on the meter will enter sleep mode when no keys are...

Страница 20: ...ta logs calibration logs and settings and return the meter to its factory default state Press the or key to scroll through the Setup Menu list until the Factory Reset item is shown Press the or key to...

Страница 21: ...s to be manually adjusted use the or key to edit the value of the pH buffer at its measured temperature To eliminate temperature errors associated with the pH electrode use an automatic temperature co...

Страница 22: ...ion The active Calibration Buffer Set is shown Rinse the pH electrode and ATC probe and place into the pH 4 01 buffer Wait for the pH value to stabilize Once the reading is stable press the Enter key...

Страница 23: ...11 12 14 40 1 69 4 03 6 84 9 07 11 99 50 1 71 4 06 6 83 9 01 11 73 60 1 72 4 09 6 84 8 96 11 50 70 1 74 4 12 6 85 8 92 11 30 80 1 77 4 16 6 86 8 89 11 13 DIN Buffer Set Temp C pH 1 09 Buffer pH 3 06...

Страница 24: ...Calibration The thermistor sensor used for automatic temperature compensation and measurement is both accurate and stable so frequent calibration is not required When an ATC probe is connected to the...

Страница 25: ...are automatically saved to the data log and exported at the selected time intervals from 5 seconds to 60 minutes the entire time the meter is in the measurement mode 5 2 pH and ORP Measurements Rinse...

Страница 26: ...Baud Rate 9600 bps Data bits 8 Parity None Stop bits 1 Flow Control None 5 5 1 Remote Commands Remote commands allow the meter to be interfaced with computer software like LIMs and HyperTerminal The r...

Страница 27: ...pH buffers and pH electrode o Replace pH buffers o Clean and condition pH electrode o Replace pH electrode as needed 6 2 Measurement Troubleshooting Reading is unstable slow to stabilize Clean electr...

Страница 28: ...t Up to 5 C up to 9 F Features Display 5 backlit LCD Time and Date Yes Read Types Continuous Auto Read Timed Timed Interval 5 seconds to 60 minutes Hold Function Yes Data Log 500 data sets with time a...

Страница 29: ...ired on a charged basis The warranty on your meter shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the...

Страница 30: ...ironment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communicat...

Страница 31: ...entsorgt werden Mehr Informationen ber die Einhaltung dieser Anweisungen durch Thermo Scientific dieVerwerter und Hinweise die Ihnen n tzlich sein k nnen die Thermo Fisher Scientific Produkte zu iden...

Страница 32: ...13 620 19 accumet ORP electrode 13 620 81 accumet meter attached electrode stand 13 637 671 accumet AB315 and AB330 meter replacement universal power cord 110 240V 13 637 010 accumet AB315 and AB330...

Отзывы: