3.2.2 System Start Up
3.2.2.1 Plug in the Power Cord, Turn on Power
After turning on the power, system displays system type, serial number (S /
N) for 5 seconds. At this point, system runs self-testing programs in the
background.
System automatically checks water pressure, determines the water supply
situation. If no water supplied, or inlet pressure is too low, monitor
displays "Low Pressure" as shown below, and system goes to standby.
3.2.2.2 Start Up the System
When the START button is pressed, system enters operational mode after
an automatic flush of the RO membrane for 180 seconds. The System
automatically detects the quality of the RO water, discharges it to the drain
if not meeting quality parameters, or sends it to the water tank if it passes
quality parameters. At system initial start-up, or after installing a new RO
membrane, allow system to run at least one hour to clean up the RO
membrane.
a. After the initial installation of a new system, it may take 2 hours or
Accu500
S/N: S0RD50503
14/10/17 14:59:59
RO: 12.0
µ
S/cm 25
UP: 1.0 M
Ω
.cm 25
LOW PRESSURE