longer before you can dispense water as system is conducting initial
RO rinsing and filling the water tank.
b. System automatically runs a rinsing process for a newly installed RO
membrane to clean up its preservative solution. You may not be able to
dispense water during rinsing. RO indicator light blinks, indicating water
quality has not reached preset quality standards. Once the new RO
membrane is thoroughly rinsed, water production will return to normal.
c. System goes to standby automatically once the water tank is full.
System displays "Tank Full". When water is being dispensed, system
will automatically restart production of water.
d. If an excessive amount of water is dispensed, water level inside the PE
tank may fall below the minimum set-point. Water dispensing is stopped
automatically. System will display "Tank Empty" on the display panel.
Once water tank is refilled sufficiently, water can be dispensed again.
3.2.3 System Degassing
1) Press the UP button to dispense UP water for 3 - 5 minutes.
2) System displays resistivity of the UP water. At initial start-up, UP light
may blink for a few seconds, indicating UP water not to preset standards.
Continue to dispense ultrapure water until the light stops blinking
and no air bubbles come out the ultrapure spout.
3) Check system leakage. Should there be a leak, reconnect tubing or parts.
4) Reinstall the system cover and tighten screws.
5) Press the UP button again to stop dispensing.
6) Install system side panels if no visible leakage from connections
inside the system.