www.eu.fishersci.com/fisherbrand
40
Attention!
DO NOT attempt to pull the tubing out from the RO housing without unlocking
the connector first or use excessive force to pull the tubing out as this will
damage the connector, rendering it useless.
5. Press down the collets at joints with water tubing and then pull tubing out
in a gentle manner.
6. Connect labeled tubing inside the system to the matching labels (RO IN,
RO OUT and RO DRAIN) on the new membrane housing
7. Install the assembled RO membrane onto the support rack.
8. Power up the system, then press START button.
9. Check system leakage. If there is a leak, reconnect the cartridge.
10. Press the START button to put system into Standby.
11. Install the system side panel.
Collet
(push in)