9.  Maintenance 
 

It is recommended to periodically inspect the unit, both visually and physically, to insure optimum 
and safe performance.  This inspection should be scheduled as a routine maintenance 
procedure, done with the unit power OFF and with the unit unplugged from the AC power source. 
 
Long exposure to acids or caustics results in corrosion of metal parts or components. Check the 
generator, converter, and cables periodically for any signs of rust or discoloration.  If 
discoloration is found, move the ultrasonic processor away from the source of the contaminant. 
 
Examine the condition of the high voltage cable that attaches the converter to the generator.  
Inspect the wire insulation for damage, such as wear, burning from hot plate contact or breakage 
from extended use or rough handling.  In general use, the cable assembly should not be used to 
carry the converter or pull it toward the user.  Make certain the cable always has slack and is 
never tensioned.  If necessary, move the generator or converter assembly closer to one another 
to accomplish this.  If this is not possible, contact your Customer Service Representative to 
obtain a longer cable.  

WARNING:  Do not use a cable with broken end connections, 

exposed wires or frayed insulation.  High voltage is present in the cable and will pose a 
shock hazard.  Do not touch the converter assembly until the power switch is off and the 
unit is unplugged.  
 
Microtip/ Probe Maintenance 

Ultrasonic processors create high intensity vibration which puts stress on the converter and horn 
assembly.   The sides and end of the probe must 

never

 be allowed to come in contact with 

anything but the solution.  When using a Microtip, the stress resulting at the point of contact with 
the vessel could cause the Microtip to fracture.   
 
Proper care of the probe is essential for dependable operation.  The intense cavitation will, after 
usage for period of time, cause the tip to erode, and the power output to decrease.  The 
smoother and shinier the tip, the more power will be transmitted into the sample. The vibrations 
may also cause the probe tip to loosen over time or the threaded connection to accumulate 
debris.  

Note:

  

A loose probe will usually generate a loud piercing or squealing sound.   

 

For that reason, 

it is recommended that a preventative maintenance schedule be adopted 

to examine the unit at regular intervals.

  The schedule should depend on frequency of use.  

Weekly maintenance schedules are recommended for units used frequently or monthly for those 
used infrequently.  The tip must be examined for excessive wear and to ensure that the 
threaded connection is clean and attached properly to the convertor.  Use a cotton swab and 
alcohol (i.e. ethanol, isopropyl, etc.) to clean the threaded mating surfaces.

    

 
When excessive wear (corrosion/pitting of the probe tip) is detected the probe should be 
replaced with a new one.

 

 

WARNING:  Never hand tighten probes or horns onto the convertor; properly tighten 
them with the appropriate Wrench Set.  

Steps for attaching and detaching microtip probes: 

 
 
 
 
 
 

Содержание FB-705

Страница 1: ...Rev 6 12 1 Sonic Dismembrator ULTRASONIC PROCESSOR Part No FB 705 OPERATION MANUAL...

Страница 2: ...Rev 6 12 2...

Страница 3: ...les of Operation 9 10 5 Description of Components Functions of Controls 11 13 6 Preparation for Use 14 7 Operating Instructions Getting Started 15 24 8 Accessories 25 26 9 Maintenance 26 28 10 Trouble...

Страница 4: ...rranty of either merchantability or fitness for a particular purpose No person or company is authorized to change modify or amend the terms of this warranty in any manner or fashion whatsoever Under n...

Страница 5: ...ity Do not place the Generator within a Fume Hood Hearing protection is highly recommended It is recommended that a sound abating enclosure or ear protection be used when operating the Ultrasonic Proc...

Страница 6: ...Rev 6 12 6 AC power Symbols Caution Risk of electric shock Hazardous voltage Caution Risk of danger Refer to User Manual...

Страница 7: ...ng 15 Amps 8 Amps Weight 16 lbs 7 3 Kg Dimensions 8 W x 15 L x 9 H 203 mm x 381 mm x 229 mm Output Voltage 1000 V rms max Output Frequency 20 KHz Converter Weight 2 lbs 900 g Dimensions 7 25 L x 2 5 D...

Страница 8: ...t be used If the 220V plug is not configured to match the wall receptacle a properly grounded universal AC socket adapter must be added Important Universal adapters do not convert voltage or frequency...

Страница 9: ...ors have a wattage display During operation the wattage displayed is the energy required to drive the radiating face of a probe at that specific amplitude setting against a specific load at that parti...

Страница 10: ...he sample larger probe diameter or higher pressure the greater the amount of power that will be delivered to the probe Setting the AMPLITUDE control to its maximum will not cause the maximum power rat...

Страница 11: ...2 diameter probe FB4220 FB 705 Front Panel FB 705 Rear Panel FUNCTIONS OF KEYS CONTROLS INDICATORS AND CONNECTORS Generator Converter Horn with replaceable tip Converter Cable Power Cord Connector Co...

Страница 12: ...creens CLR key Deletes selected parameter entry RUN key To proceed to program run screen SAVE key Stores a program to the selected memory number Up to 10 programs 0 9 can be stored START STOP key Star...

Страница 13: ...tch Connector Connects to the footswitch cable Temperature Probe Connector Connects to the Temperature monitoring probe or thermocouple Converter Cable Connector Output Connects to the converter Power...

Страница 14: ...E DAMAGED EQUIPMENT Retain all packing materials for future shipment ELECTRICAL REQUIREMENTS The ultrasonic processor requires a fused single phase 3 terminal grounding type electrical outlet For powe...

Страница 15: ...face tension liquids ALWAYS use a solid probe Set up 1 Connect the power cord into the receptacle on the rear of the ultrasonic processor 2 Make sure the unit is switched off Plug the electrical line...

Страница 16: ...fittings on top of the converter to a source of dry compressed air This will cool the convertor so that it can function normally under load for an extended duration without becoming too hot Operation...

Страница 17: ...ed b Program Mode Allows the user to create a program with specific On Off times and output setting The ability to save up to 10 programs and sequence several individual programs together can also be...

Страница 18: ...e monitoring A check mark will appear indicating that temperature monitoring is activated Select the appropriate Amplitude intensity setting for your sample by touching the Up or Down arrow Touch Star...

Страница 19: ...different programs Select Modify Sequence 1 2 Select a sequence of programs for the unit to run in succession A maximum of 6 programs can be sequenced at one time Option Screen The Option menu enable...

Страница 20: ...the unit will run continuously without pulsing Process Time Pulse On and Pulse Off times are measured in Hours Minutes Seconds format 6 Activate Temperature Monitoring if necessary 7 Program a Shutdow...

Страница 21: ...s The Temperature Monitoring and Shutdown Temperature option have been selected during this example program Program Progress is displayed by a bar graph This displays the progress of the current progr...

Страница 22: ...rder in the PGM fields 3 If a delay or rest time between programs is desired touch the Delay After field Enter the appropriate time frame 4 Unused PGM fields should have a dash Select unsused programs...

Страница 23: ...quid sample If the initial sample temperature is above the set point sonication will not turn on During sonication if the temperature reaches the Shutdown value the unit will pause sonication The unit...

Страница 24: ...ns Viscous solutions and highly concentrated liquids can be difficult to sonicate If the liquid is so thick that it will not pour or circulate easily or if 25 micron glass powder will not settle to th...

Страница 25: ...3 4 probe with replaceable tip 25 500ml Catalog FB4207 Note 3 4 solid probe is also available Catalog FB4208 Solid probes are required for low surface tension liquids i e organic solvents F 1 probe w...

Страница 26: ...Microtip Probe Maintenance Ultrasonic processors create high intensity vibration which puts stress on the converter and horn assembly The sides and end of the probe must never be allowed to come in c...

Страница 27: ...alcohol to dry completely 4 Reattach probe to convertor Screw the probe back onto the convertor and tighten with the wrench set provided 5 The tips on replaceable tip probes can be removed for cleani...

Страница 28: ...tions The generator and converter may be cleaned using an acid free cleaning solution i e glass cleaner Probes should be cleaned using isopropyl alcohol Probes are made from titanium and can be autocl...

Страница 29: ...life A fuse s has failed OVERLOAD CONDITION If the Ultrasonic Processor stops working and an OVERLOAD indication is displayed on the screen check for possible causes as outlined in the above paragrap...

Страница 30: ...should be exercised to provide adequate packing to insure against possible damage in shipment The Ultrasonic Processor should be sent to the Service Department with all transportation charges prepaid...

Страница 31: ...h or exposed to radiological biological or chemical agents and has been decontaminated rendering it safer for handling use or disposal Authorization By accepting authorization to return the equipment...

Страница 32: ...eater than 10 minutes of continuous processing there is potential for the probe converter assembly to increase in temperature The general rule is that if the converter is warm to the touch it should b...

Страница 33: ...a volume larger than indicated on the chart unless the application is reviewed and approved by a Qsonica representative While there is no absolute sample volume range for any probe horn below is a ge...

Страница 34: ...movement up and down is called its amplitude The amplitude is adjustable Each probe has a maximum amplitude value For example with a diameter probe at setting 100 the probe will achieve an amplitude o...

Отзывы: