Rev. 6-12
18
The following screen displays the
Manual Mode
.
A
. Microtip Mode – Indicates that the unit is set for use with microtips only. This mode should
not be used with standard probes or horns (1/2” or larger). See page 15 for information on
selecting YES or NO at the “Are you using a Microtip screen”.
B
. Amplitude (intensity) setting - Output amplitude may be set between 1-100%.
C
. Power displayed in Watts.
D
. Energy displayed in Joules.
E
. Elapsed Time – Total time of active sonication.
F
. Temperature probe measurement (note: if “OPN” appears, it indicates temperature
monitoring has been activated but the probe is not connected).
G
. Temperature probe On/Off indicator. Touch the box to activate temperature monitoring. A
check mark will appear indicating that temperature monitoring is activated.
Select the appropriate Amplitude (intensity setting) for your sample by touching the Up or Down
arrow. Touch Start to begin sonication. Touch Pause to pause sonication.
Manual and Program modes both show the Start, Pause and Exit buttons. After touching the
Start button and activating sonication, the Start button becomes a Stop button. After touching
the Pause button, the Pause button becomes the Resume button.
If sonication is stopped, Energy and Elapsed Time values remain on display. If the Start button
is touched again, the values are reset. If sonication is Paused and Resumed, the Energy and
Elapsed time data will resume counting from the point at which it was paused.
The Temperature Probe option can be selected if you wish to monitor the temperature of the
sample being processed.