The following Option Screen can be accessed from either Menu Screen.
The Option menu enables the selection of
Temperature units
and
Footswitch
operational
mode
.
Option Screen
Temperature Units
– A temperature probe (Part# 4102 or 4103) is required for use of this
option.
The Temperature Probe option can be selected if you wish to monitor the temperature of the
sample being processed.
The temperature probe must be plugged into the back of the ultrasonic processor and the probe
tip must be in the liquid sample. If the initial sample temperature is above the set point,
sonication will not turn on. During sonication, if the temperature reaches the Shutdown value,
the unit will pause sonication. The unit will remain paused until the temperature drops below the
pre-set temperature limit, at which time sonication can be manually resumed.
Footswitch Operation
– A footswitch (# FS-3) is required for use of this option. After setting up
a program, the footswitch can be used to remotely activate the unit.
Exit
– The Exit button will save changes and return to the previous screen.
Techniques for Optimizing Results
Probe Depth
Immerse the probe tip 1.5 times the tip diameter into the solution, without touching the bottom.