What do I do if...?
The behrotest
®
TRS 300 control unit has an internal fault scanning system. This is capable of detecting
five different error situations. „Error“ will then appear on a red background, together with a Code No.,
on the display. An alarm signal (bleep tones) is also heard. The numbers have the following meanings:
Code Meaning
Remedy
002
The temperature sensor is not
connected or not firmly inserted
Switch the unit off. Connect the temperature sensor of the
digester to the „Temp Probe Pt 100“ socket (Page 7, 2.2).
Ensure firm seating of the diode connector. Secure the
connector using the union nut. Only then switch the unit
on again.
003
Short-circuit in temperature
sensor
The digester must be repaired or replaced. Switch
the digester and the TRS 300 off and inform the behr
after-sales service.