5.1.19
Program 18: Clear the user standardization
Press enter to select to clear the calibration and return to
measure screen or press
setup
again to skip the option.
clear
n
n
press to select options
press to accept
setup
enter
µ
S
F
F
F
F
Depends on the measure mode you are in, this setup page
clears only the particular mode calibrated value.
Suppose you access the setup pages from TDS measure mode
and you choose to clear the standardized buffer then the meter
only resets the TDS standardization. Conductivity/ Resistivity
standardization if any, will remain unchanged.
F
F
F
F
When accessing the cell constant, temperature coefficient, TDS
factor, temperature unit selection and stable ON/OFF setup
pages, meter displays the last selected choice and pressing
ÔsetupÕ key, takes you to other options.