After adjusted std solution value (which will be displayed at the
upper display), press
std
to initiate the standardization and on
successful completion meter returns to measurement mode.
S TA BLE
n
n
press to s ele ct options
press to standardize
se tup
st d
STD Solution
mS
F
F
F
F
Standardization window provided is ±40% of current reading. If
the standard value set is not within the window then meter
displays ÒElectrode ErrorÓ message after initiate the
standardization.
Press
enter
to return to measurement mode. In this case meter
will not accept the standardization,
If the standardization is successful, then meter displays buffer
icon in the middle part of the display in the measurement mode.