FISCHER Mess- und Regltechnik GmbH
Commissioning | 3
BA_EN_MS13
19 / 32
Parameter name
Description
Value range
∙∙∙
Value pair30
Value pair 30
The display range is defined with the parameters
MB decimal pl.
,
MB start
and
MB end
. The user can select the configuration freely.
Using the parameter
MB decimal pl.
, it is possible to select between a 5 or 6-
digit presentation. The resolution is not increased. Only an extra zero or two
zeros are added. This serves the correct display of larger values. The measur-
ing range must be positive for the 6 digit presentation.
The
MB unit
gives the user the option of defining a completely independent
unit. Letters, numbers or special characters can be used. The unit can be max.
4 characters long.
If the function TABLE is selected, then it is also necessary to state the
No. of
pairs
. It is defined here how many pairs of values (support points) are used in
the table. A table is made up of at least 3, max. 30 support points.
NOTICE
Number of value pairs
If the number of value pairs is changed, the table is initialised again and the ex-
isting values are deleted.
Value pair
2
+14,6 mbar
+8,6 %
+0,0 ... +100,0 mbar
1
2
1 input mark (value flashes)
2 allowed range of values
Illustration 17:
Value pair
Start
End
Measuring range
Output / display [%]
100
0
0.0
100.0
Value pair 2
Illustration 18:
Table function (example)
The individual value pairs can be seen and changed with the
Value pair1
to
Value pair30
parameters. A value pair comprises a measured value (left side)
and a display value (right side). The measured value must lie within the measur-
ing range and the display value must lie within the defined "free unit". The re-
spective limits are shown during input. The table must contain either increasing
or decreasing values. the table must contain either continuously increasing or
continuously failing values. A change from an increasing to a decreasing char-
acteristic curve within a support point table is not allowed.